6.15.2 Set the pipetting properties
Abb. 6-10: Pi petting Settings
Fig. 6-10: Pipetting Settings
6.15.3 Liquid Types – Setting the liquid properties
The liquid properties allow adjusting the pipetting to liquids with densities or vapour pressures that differ
sigificantly from those of water. The preset Standard is used for verifying the device and is suitable for
water and diluted aqueous solutions.
Parameter Value Pipette tips Default settings
Reverse stroke (μL) Setting the additional volume for the
liquid aspiration in μL.
50 μL 3 μL
300 μL 12 μL
Remaining stroke (μL) Setting the residual volume in μL for
liquid aspiration in Multidispense mode.
50 μL 4 μL
300 μL 12 μL
Blow-out (uL) Setting the air volume in μL for
dispensing the residual liquid.
50 μL 15 μL
300 μL 20 μL
Piston Pause (s) Setting the delay time between pipetting
operations in seconds.
–1 s
Auto Delay (s) Setting the delay time for automatic
liquid dispensing in seconds
(Multidispense mode).
–1 s