Operation
epMotion
®
96
English (EN)
48
6.13 Mode Multiaspirate – Multiple aspiration of liquid
Aspirating liquid in several steps and dispensing liquid in one step.
Abb. 6-6: Mode Multiaspirate
Fig. 6-6: Mode Multiaspirate
6.13.1 Aspirating liquid
Prerequisites
• Source vessel is available.
1. Put the source vessel onto the lifting table and move the lifting table into working position.
2. Set the aspiration volumes and the speed levels.
3. Set the number of aspiration steps.
4. Press Aspirate.
5. Change or reposition the source vessel.
Parameter Value Range of values
Speed aspirate Setting the speed of the liquid aspiration. 1 – 9
Speed dispense Setting the speed of the liquid dispensing. 1 – 9
Aspiration Volume Setting the aspiration volume in μL. 0.5 – 300 (increment of 0.1)
Steps Setting the number of aspiration steps. 1 – 99