Operation
epMotion
®
96
English (EN)
36
6. Tighten the locking handle on the lifting table.
7. Press Aspirate.
Liquid is aspirated.
Destination vessel can be filled.
6.6 Dispensing liquid
Prerequisites
The lifting table is in the basic position.
Destination vessel is available.
1. Put the destination vessel onto the lifting table.
2. Set the parameters.
3. Loosen the locking handle on the lifting table.
4. Move the lifting table into the working position until the openings of the pipette tips are positioned
below the rim of the vessel.
5. Tighten the locking handle on the lifting table.
6. Press the Dispense button.
7. Loosen the locking handle.
8. Move the lifting table upwards until the pipette tips are immersed in the liquid.
Adherent liquid droplets are stripped off.
9. Move the lifting table into basic position.
6.6.1 Blow out – Dispensing residual liquid
Prerequisites
Blow out button is active (is displayed instead of Dispense).
Destination vessel is available.
1. Move the lifting table to the operating position and fix it using the locking handle.
2. Press the Blow out button.
3. Strip off liquid droplets at the pipette tips.
4. Move the lifting table into basic position.
5. Confirm the query with OK.
Pipette tips are emptied.
6.6.2 Empty – Completely dispensing liquid
1. Move the lifting table to the operating position and fix it using the locking handle.
2. Press the Empty button.
Pipette tips are emptied entirely.
Page view 35