Operation
epMotion
®
96
English (EN)
34
Prerequisites
The lifting table is in the basic position.
Pipette tips (epT.I.P.S Motion Reloads) are available.
6.4 Moving the lifting table
6.4.1 Moving the lifting table into working position
Prerequisites
The pipette tips are attached.
1. Pull the lever of the pipetting device to the front.
2. Pull out the loading frame.
3. Insert the tray with pipette tips into the loading
frame.
The notch on the tray must point in the direction
of the loading frame handle.
4. Insert the loading frame into the pipetting device.
5. Push the lever of the pipetting device backwards.
The pipette tips are attached and locked.
The status bar of the software indicates the
volume of the pipette tips.
1. Loosen the locking handle.
2. Move the lifting table upwards into working
position.
3. Tighten the locking handle.
Liquid can be aspired.
Liquid can be dispensed.
Stop can be set.