Operation
epMotion
®
96
English (EN)
6.12.1 Aspirating liquid
Prerequisites
Source vessel is available.
1. Set the speed levels.
2. Set the liquid volume for the dilution solution.
3. Set the volume for the air gap.
4. Set the liquid volume for the concentrate.
5. Put the source vessel with the dilution solution onto the lifting table and move the lifting table into
working position.
6. Press Aspirate.
Dilution solution is aspirated.
7. Move the lifting table into basic position.
8. Press Aspirate.
Air gap is aspirated.
9. Put the source vessel with the concentrate onto the lifting table and move the lifting table into working
position.
10.Press Aspirate.
Concentrate is aspirated.
Target vessel can be filled.
6.12.2 Diluting and mixing liquids
Prerequisites
Liquids and air gaps are aspirated.
Destination vessel is available.
1. Put the destination vessel onto the lifting table and move the lifting table into working position.
2. Set the number of mixing cycles.
3. Press Dispense.
Liquids are dispensed and mixed automatically.