6.12 Mode Dilute and Mix – Mixing liquid
The application is suitable for diluting samples and reagents using an appropriate dilution solution.
Aspirate the concentrated liquid, aspirate diluent and mix while dispensing.
Abb. 6-5: Mode Dilute and Mix
Fig. 6-5: Mode Dilute and Mix
Parameter Value Range of values
Speed aspirate Setting the speed of the liquid aspiration. 1 – 9
Speed dispense Setting the speed of the liquid dispensing. 1 – 9
Diluent Vol. Setting the aspiration volume for the diluent in
μL.
0.5 – 300 (increment of 0.1)
Air Gap Vol. Setting the volume of the air gap in μL. 0.5 – 300 (increment of 0.1)
Concentrate Vol. Setting the aspiration volume of the concentrate
in μL.
0.5 – 300 (increment of 0.1)
Mix vol. Setting the mixing volume in μL. 0.5 – 300 (increment of 0.1)
Mix Cycles Setting the number of mixing cycles. 1 – 99