Eppendorf 5430 R Centrifuge User Manual Page 60

  • Download
  • Add to my manuals
  • Print
  • Page
    / 130
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 59
Technical data
Centrifuge 5430/5430 R
English (EN)
60
10.3 Weight/dimensions
Centrifuge 5430
Centrifuge 5430 R
*) The noise level was measured frontally in a sound measuring room with accuracy class 1, 1 m from the
device and at lab bench height.
10.4 Application parameters
Dimensions Width: 335 mm (11.2 in.)
Depth: 415 mm (16.3 in.)
Height: 250 mm (9.84 in.)
Weight without rotor 29 kg (63.9 lbs.)
Noise level < 60 dB(A) *
Dimensions Width: 380 mm (15.0 in.)
Depth: 640 mm (25.2 in.)
Height: 296 mm (11.7 in.)
Weight without rotor 56 kg (123.5 lbs.)
Noise level < 60 dB(A) *
Run time 30 s to 99:59 h, infinity ( )
Adjustable to 10 min in 0.5 min increments, then increments of
1 min
Temperature of 5430 R -11°C to 40°C
Relative centrifugal force (or rcf) 1 to 30.130 × g
Adjustable in increments of 10 × g up to 3,000 × g, then in
increments of 100 × g
Speed 100 to 17,500 rpm
Adjustable in increments of 10 rpm up to 5,000 rpm, then in
increments of 100 rpm
Maximum load 48 tubes with 2.0 mL each or 6 conical tubes with 50 mL each
Maximum kinetic energy 10,000 nm
Compulsory test log book No
Permitted density of the material for
centrifuging at max. g-force/rpm and
max. load
1.2 g/mL
Page view 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 129 130

Comments to this Manuals