InjectMan® NI 2
Operating Manual
T_R_InjectMan_NI2.fm Seite 1 Donnerstag, 6. Dezember 2012 6:03 18
Page view 0
1 2 3 4 5 6 ... 59 60

Summary of Contents

Page 1 - InjectMan® NI 2

InjectMan® NI 2 Operating ManualT_R_InjectMan_NI2.fm Seite 1 Donnerstag, 6. Dezember 2012 6:03 18

Page 2

63On the A module is the tool mount for the probe holders. Tools with a diameter, or width, of 4 mm and weighing max. 100 g can be inserted, e.g. cap

Page 3 - Contents

64 3.3.1 Tool movement Moving the dynamic joysticks moves the capillaries in one plane. The speed of the motion is determined by the speed of the de

Page 4

65 Axial Activate or disable axial movement. Pos 1 / Pos 2 Approach a stored position.Store current position by pressing and holding down the key.Cle

Page 5 - 1 Summary

66 Turn wheel Set the speed in the current speed mode. Display of selected speed on-screen. 3.3.3 Display In normal operation the display shows the

Page 6

67– The packaging and the transport guards should be kept for use if the device is transported again later! – The control board must not be carried o

Page 7

68 4.2.1 Components of the module unit Fig. 3: Module unit, assembled (here right side mounted)1 Motor unit for horizontal movements (X/Y module)

Page 8

69The hole spacing is 5°. By moving the mounting point of the angle guide 5° upwards, the adjustable angle changes each time to 50° to 30°. By moving

Page 9

70 4.2.4 Angle adjuster Set the desired angle. Mounting the angle guide at the centre hole and adjusting the A module to 0° corresponds to an angle

Page 10 - 3 Description

71For right side mounting, the hinge must be at the right outer (not visible here, on the rear).The module unit is mounted from above onto the hinged

Page 11

72The hinged bracket is opened up and joined to the motor module by its inner inclined surface by means of the M5x16 screw. Close the hinged bracket.

Page 12

2Bedienungsanleitung. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Operating Ma

Page 13

73 4.3.1 Right and left side mounting The module unit mounting is dependent on the chosen side of the microscope.The module unit can be mounted on t

Page 14 - 4 Installation

74According to the mounting mode on the left side of the adapter. Note : In the "Install" menu (section 5.4, subsection 5.4.2.6) the parame

Page 15

75To avoid having to reposition every time a capillary is changed, the positioning aid can be mounted on the universal capillary holder so it fits in

Page 16

76 4.4.1 Transmission cable 5181/5246 For optimum use of the INJECT function (see section 5.1), the InjectMan ® NI 2 micromanipulator and the micro

Page 17

77Section 5.1 provides a brief summary of how the commands can be used for the injection procedure. Section 5.2 provides additional information on th

Page 18

78The new speed is shown on the display as a modified block diagram.If the capillary touches the medium surface for example (visually monitored), swit

Page 19

79 Changing the limit As a result of moving the petri dish, the limit may need to be adjusted. This is done using the arrow keys to the left and righ

Page 20

80– Carefully slacken knurled screw of tool mount and positioning aid a little way. Carefully bring the capillary holder with the capillary into the

Page 21

81Or the capillary is brought into the focal plane like the cell. – Press the "Limit" key.The defined limit is indicated on the display.To c

Page 22

82 5.2.3 Injection Pressing the joystick button activates the Inject function. The capillary moves automatically from its current position, the sear

Page 23

561 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 582 Intended use

Page 24

83– Synchronisation Move: No pressure build-up occurs. If the joystick button is held down the capillary stops as shown in Fig. 6.3 and a separate in

Page 25

84 5.2.5 Field of movement To utilise the benefits of axial injection to the full, only one motor (A module) is used. This permits less harmful injec

Page 26 - or joystick

85The keypad can be used to make settings or initiate movements directly. Frequently used settings can be entered by means of key combinations. The L

Page 27

86 5.3.2 Setting the speed The speed can be set in three ways:Directly using the turn wheel, by pressing and holding down the key, or by way of the

Page 28

87 5.3.3 Home The tools moves axially out of the working radius.The function is intended for changing capillaries, or for quick removal of capillari

Page 29 - Z-Limit

88 5.3.5 Pos 1 / Pos 2Approach a stored position. Store current position by pressing and holding down the key (acoustic acknowledgement). Clear

Page 30

895.3.8 MenuToggles between menu and back. Parameters can be selected or changed in the menu using the joystick. For a detailed description see secti

Page 31

905.3.11 Injection speedBy pressing and holding down the "Limit" key the injection speed can be adjusted. The current selected speed is ind

Page 32

91Moving joystick to the right moves the cursor into the submenu. The left hand column shows the possible parameters, the right hand column the associ

Page 33 - 250 µm/s

92To switch to other main menu options press the joystick button again or move the joystick to the left.Alternatively, the or keycan be used.To swi

Page 34

575.4 Parameter settings via menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 935.4.1 Meanings of param

Page 35

935.4.1 Meanings of parameters in menu5.4.1.1 Main menu optionsSpeed: Includes all speed parameters.Inject: Includes all the parameters relevant t

Page 36

945.4.2.2 Inject InjSpeed: Speed of (axial) injection movement. Adjustable from 7,500 to 5 µm/s in 5-µm/s increments.Synchron: Synchronisation of co

Page 37

955.4.2.3 StepInjThis special parameter enables the tool to be move forward automatically from its current position by a specific distance (D) at a sp

Page 38

965.4.2.5 PositionPosSpeed: Sets the speed at which the stored position is approached.Adjustable from 7,500 to 5 µm/s in 5-µm/s increments.ClearAll:

Page 39 - –Submenu –

975.4.2.7 FunctionZeroCord: This function sets the coordinates of the X, Y and Z axes to zero. CentrMot: This function sets the motors to a mid pos

Page 40

985.5.2 Software version pollingAs an additional software version polling facility (also visible during the power-up routine on the display) the vers

Page 41

99Before cleaning the InjectMan ® NI 2, disconnect the plug.Ensure that no fluids enter into the inside of the micromanipulator in order to avoid sho

Page 42

100The following error situations are conceivable: Remedy – Motor axes move in wrong direction or do not match joystick movement.– Check installation

Page 43

101The following warnings or error messages may be displayed: Remedy – Warning 001 Check Injector and cable: The injector is not ready or there is no

Page 44

102– Error 044 – 046In self-test wrong values defined for displayed axis.– Repeat self-test, reset, general reset. If error recurs: Service– Error 06

Page 45

58The Eppendorf InjectMan ® NI 2 micromanipulator is specially designed as a micromanipulator for use in microinjection. Its optimum functionality a

Page 46 - 6 Cleaning and service

103 Power supply Input: 120 – 240 V AC; 50 – 60 Hz; < 50 WOutput: 15 – 18 V DC; 25 W Control board 15 – 18 V DC; 25 WContamination level: II O

Page 47 - 7 Troubleshooting

104 Order no. InjectMan ® NI 2, Europe 5181 000.017InjectMan ® NI 2, USA 5181 000.025InjectMan ® NI 2, UK/Hongkong 5181 000.033InjectMan ® NI 2,

Page 48

105 Accessories (non-standard) Order no. Universal capillary holder for accommodating Femtotips ® , Femtotips ® II or grip heads5176 190.002Grip he

Page 49

106 Order no. FemtoJet ® expressProgrammable microinjector incl. connection for external pressure supplyVersion Europe 5248 000.017Version USA and C

Page 50 - 8 Technical data

107 Order no. InjectMan ® NI 2, USA 920000029InjectMan ® NI 2, international 5181 000.017InjectMan ® NI 2, UK/Hongkong 5181 000.033InjectMan ® N

Page 51 - 9 Ordering information

108 Accessories (non-standard) Order no. Universal capillary holder for accommodating Femtotips ® , Femtotips ® II or grip heads920007392Grip head

Page 52

109 Order no. FemtoJet ® expressProgrammable microinjector incl. connection for external pressure supplyVersion USA and Canada 5248 000.050Version E

Page 53

110From the Install submenu up to 4 Users (users/applications) can be activated. The Users can be assigned dedicated parameters for speed, injection

Page 54

03_Fehler_sp.fm Seite 144 Donnerstag, 6. Dezember 2012 5:41 17

Page 55

T_R_InjectMan_NI2.fm Seite 3 Donnerstag, 6. Dezember 2012 6:03 18

Page 56

59The InjectMan NI 2 enables an electronically controlled axial injection movement. It is designed for controlled, gentle, semi-automatic microinject

Page 57 - A Appendix – MultiUser

Your local distributor: www.eppendorf.com/worldwideEppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 53801-0 · Fax: +49 40 538 01-556 · E-mail: epp

Page 58

60 • Caution when mounting tools!When the tool is mounted, the motors are swung out forwards and the capillaries face the front. Handle with care! •

Page 59

61Before checking that the device has been delivered complete and installing it, please read the following section thoroughly. • The control board mu

Page 60

62The Eppendorf InjectMan ® NI 2 micromanipulator is made up of three components:– Control board– Module unit– Mains power supply unitFig. 1: Injec