Eppendorf 5430 R Centrifuge User Manual Page 59

  • Download
  • Add to my manuals
  • Print
  • Page
    / 130
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 58
Technical data
Centrifuge 5430/5430 R
English (EN)
59
10 Technical data
10.1 Power supply
Centrifuge 5430
Centrifuge 5430 R
10.2 Ambient conditions
Mains connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption 3 A (230 V)
6 A (120 V)
7 A (100 V)
Power consumption Maximum 475 W
EMC: interference emission (radio
interference)
EN 61326 - category B
EMC: noise immunity EN 61326
Overvoltage category II
Protection class 1
Fuses 4.0 AT (230 V)
8.0 AT (120 V/100 V)
Mains connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption 6 A (230 V)
12 A (120 V/ 100 V)
Power consumption Maximum 1050 W
EMC: Interference emission (radio
interference)
EN 61326 – Class B (230 V/120 V)
EN 61326 – Class A (100 V)
EMC: Noise immunity EN 61326
Overvoltage category II
Protection class 1
Fuses Thermal overcurrent protective switch 7 A (230 V)
Thermal overcurrent protective switch 15 A (120 V/100 V)
Ambience For indoor use only.
Ambient temperature Centrifuge 5430: 4 to 40°C
Centrifuge 5430 R: 10 to 40°C
Maximum relative humidity 75 %, non-condensing humidity
Atmospheric pressure Use up to an altitude of 2000 m above MSL.
Degree of pollution 2
Page view 58