Operating manualOperating manualEppendorf Research® plusOperating manualResearch plus - Operating Manual - en
3 OperationEppendorf Research® plus — Operating manual10 3 Operation3.1 Setting the volume (only Research plus variable)3.2 Using pipette tips The
3 Operation11Eppendorf Research® plus — Operating manual 3.3 Aspirating liquid3.4 Dispensing liquid Pipette tips are for single use only.1. The liq
4 TroubleshootingEppendorf Research® plus — Operating manual12 4 Troubleshooting4.1 Error searchSymptom Possible cause SolutionLiquid is dripping f
4 Troubleshooting13Eppendorf Research® plus — Operating manual To avoid dispensing errors, check the precision and accuracy of the Research plus at
5 MaintenanceEppendorf Research® plus — Operating manual14 5 Maintenanc e• All single and multi-channel lower parts are wear parts. Clean them afte
5 Maintenance15Eppendorf Research® plus — Operating manual 5.2 Sterilizing or disinfecting the pipette5.2.1 AutoclavingAll Research plus pipettes a
5 Maintenance16Eppendorf Research® plus — Operating manual 5.2.2 Disinfection Carefully wipe off the outer surfaces with disinfectant, DNA/RNA dec
5 Maintenance17Eppendorf Research® plus — Operating manual 5.4 Research plus disassembly and assembly5.4.1 Single-channel up to 1 000 μLRemoving th
5 Maintenance18Eppendorf Research® plus — Operating manual Opening and assembling the lower part (≤1 000 μL)Hint!To avoid confusion of parts, only
5 Maintenance19Eppendorf Research® plus — Operating manual 5.4.2 5 mL and 10 mL single-channelReplace the protection filter in the tip cone after e
Research plus - Operating Manual - en-02/0210 Copyright© 2010 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior
5 Maintenance20Eppendorf Research® plus — Operating manual Before assembling the pipette, first install the piston mount, the piston spring and the
5 Maintenance21Eppendorf Research® plus — Operating manual Removing and installing channelsChannels must only be removed and installed if the multi
5 Maintenance22Eppendorf Research® plus — Operating manual 5.5 Adjusting pipettes5.5.1 General notes on user and factory adjustment settingsThe Res
5 Maintenance23Eppendorf Research® plus — Operating manual The density of distilled water is approx. 0.9982 mg/L at 20°C and 0.9965 mg/L at 27°C.
5 Maintenance24Eppendorf Research® plus — Operating manual 5.5.2 Changing the user adjustmentIf the adjustment is changed, the volume changes by a
5 Maintenance25Eppendorf Research® plus — Operating manual 5.5.3 Changing the factory adjustmentAuxiliary equipment • Supplied safety plug tool• Su
5 Maintenance26Eppendorf Research® plus — Operating manual 2. Mark the volume setting ring and the ejector with a common mark with a pen. This mark
5 Maintenance27Eppendorf Research® plus — Operating manual 5. Turn the volume setting ring slightly to change the volume. Proceed as shown in the f
5 Maintenance28Eppendorf Research® plus — Operating manual This results in the following approximate volume changes:Single-channelNominal volumeCol
5 Maintenance29Eppendorf Research® plus — Operating manual The values mentioned are theoretical values and are for orientation purposes only. The v
Table of contentsEppendorf Research® plus — Operating manual3 Table of contents1 Product description . . . . . . . . . . . . . . . . . . . . . . .
5 Maintenance30Eppendorf Research® plus — Operating manual 9. Document the changes made and the measurements conducted. Remove the mark on the volu
5 Maintenance31Eppendorf Research® plus — Operating manual 5.5.5 Error limits in accordance with ISO 8655-2Error limits in accordance with ISO 8655
5 Maintenance32Eppendorf Research® plus — Operating manual 30 - 300 LIncrement:0.02 Lorange20 - 300 L55 mm30 L ± 13.0 ± 3.9 ± 5.0 ± 1.5150 L ±
5 Maintenance33Eppendorf Research® plus — Operating manual Error limits in accordance with ISO 8655-2 - Research plus variable multi-channel Model
5 Maintenance34Eppendorf Research® plus — Operating manual 5.6 Decontamination before shipmentIf you wish to return the pipette to Eppendorf AG or
6 Technical data35Eppendorf Research® plus — Operating manual 6 Technical data6.1 Research plus single-channel variableResearch plus variable singl
6 Technical data36Eppendorf Research® plus — Operating manual 30 - 300 LIncrement:0.2 Lorange20 - 300 L55 mm30 L ± 2.5 ± 0.75 ± 0.7 ± 0.21 1
6 Technical data37Eppendorf Research® plus — Operating manual 6.2 Research plus multi-channel variableResearch plus variable multi-channelModel Tes
6 Technical data38Eppendorf Research® plus — Operating manual 6.3 Research plus fixTest conditions according to ISO 8655-1, 8655-2 and 8655-6 for p
6 Technical data39Eppendorf Research® plus — Operating manual 6.4 Ambient conditionsTechnical specifications subject to change.Temperature range Re
1 Product descriptionEppendorf Research® plus — Operating manual4 1 Product desc ription1.1 FeaturesThe Research plus pipettes are piston-stroke pi
7 Ordering informationEppendorf Research® plus — Operating manual40 7 Ordering informationHint!The ordering information can be found on the Researc
Eppendorf offices Eppendorf officesAUSTRALIA & NEW ZEALANDEppendorf South Pacific Pty. Ltd.Phone: +61 2 9889 5000Fax: +61 2 9889 5111E-mail: Info@
Your local distributor: www.eppendorf.com/worldwideEppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 538 01-0 · Fax: +49 40 538 01-556E-Mail: eppe
1 Product description5Eppendorf Research® plus — Operating manual 1.3 Main illustration1.3.1 Research plus single-channel1 Control buttonThe cont
1 Product description6Eppendorf Research® plus — Operating manual 1.3.2 Research plus multi-channelFor an explanation of the upper part of the pipe
1 Product description7Eppendorf Research® plus — Operating manual 1.4 MaterialsThe user-accessible components of the Research plus are made of the
2 SafetyEppendorf Research® plus — Operating manual8 2Safety2.1 Intended useThe Research plus is a lab device intended for dispensing liquids in th
2 Safety9Eppendorf Research® plus — Operating manual NOTICE!NOTICE! Carry-over, contamination and incorrect dispensing results due to the incorrect