Eppendorf Research
fix · variable · multi
Instruction Manual · Bedienungsanleitung · Mode demploi
Instruzioni dimpiego · Manual de Instrucciones
®
1_Research_Titel.fm Seite 1 Donnerstag, 17. April 2008 12:31 12
Page view 0
1 2 3 4 5 6 ... 41 42

Summary of Contents

Page 1 - Eppendorf Research

Eppendorf Research fix · variable · multiInstruction Manual · Bedienungsanleitung · Mode d’emploiInstruzioni d’impiego · Manual de Instrucciones®1_Res

Page 2

55 4.1 Regolazione del volume (eccetto il modello a volume fisso) Girando la ghiera di regolazione (Fig. 2-2) è possibile regolare il volume in conti

Page 3 - Research – variable / fix

56 Avvertenza: se il puntale di una pipetta Research variabile da 500–5000 µL e 1–10 mL viene estratto troppo repentinamente dal liquido, le forze c

Page 4

57Il numero di serie è riportato sulla ghiera di regolazione del pulsante di comando. 5.1 ControlloResearch monocanale L'acqua bidistillata vie

Page 5 - Research – Parte A – Indice

58– Temperatura della pipetta?– Temperatura dell'aria?– Conversione da mg in µL?– La densità del campione è diversa da quella dell'acqua?–

Page 6

59d Per ricalibrare successivamente la regolazione sul volume indicato occorre inserire orizzontalmente attraverso il sigillo di calibrazione il lato

Page 7 - 2 Dati tecnici

60Poiché la ricalibrazione ha effetto sull'intero range di misurazione, occorre assolutamente controllare anche gli altri volumi della pipetta r

Page 8

61Per mantenere inalterata la qualità delle pipette si consiglia di utilizzare i rispettivi componenti della stessa generazione(PhysioCare Concept +

Page 9

62 Variante 1–10 mL (Fig. 1-14)Sfilare il filtro dal cono di lavoro verso il basso. Durante la sterilizzazione in autoclave, il filtro si rigonfia legge

Page 10

63 Anomalia Causa Rimedio Goccioline sulla parete interna del puntale per pipette– Umidificazione irregolare della parete plastica.– puntale irregol

Page 11

64 Dubbi sulla correttezza dei dati di dosaggio Per evitare errori di dosaggio, è necessario verificare periodicamente precisione e correttezza di Res

Page 12

Research variable 0.1 – 2.5 µL 0.5 – 10 µL 2 – 20 µL 10 – 100 µL 20 – 200 µLResearch fix – 20 µL – 100 µLS

Page 13

80 I. Single-channel model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 A Disassembly and assembly . . . . .

Page 14

81 Research – Part B – Maintenance For information on replacing pistons and seals as well as on disassembling and assembling the different members of

Page 15

82Please see page 3 (Fig. 1) of this manual.A general description of disassembly and assembly for all single-channel models is followed by instructio

Page 16

83For fixed- and adjustable-volume (variable) pipettes up to 100 µL (see Fig. 1 / point 3):– Disassemble the pipette.– Using side B of the wrench, u

Page 17

84Fixed- and adjustable-volume (variable) pipettes up to 1000 µL (see Fig. 1 / point 5):– Disassemble the pipette.– Undo the screw (6) using side C of

Page 18 - 7 Ricerca dei guasti

85– Pull the locking mechanism (5a), which is located on the cylinder attachment (5), off upwards.– Unscrew cylinder (11) from the cylinder attachment

Page 19

86 A Replacing the O-rings – Hold down the ejection button on the side of the pipette and unscrew the lower part counterclockwise from the grip.– Us

Page 20 - Research – Part B – Contents

87– Please see page 115 of this manual.– Pull the housing off the cover plate.– Hold the lower part firmly, as shown in Fig. 7.– Press side B of the t

Page 21

88 30 – 300 µL – Proceed as with the other sizes, The press piece is discarded.– Using side C of the tool, pull out the spring, spring plate and seal

Page 22

89Please see page 3 (Fig. 1) of this manual.In order to ensure the consistent quality of the pipettes, we recommend using the corresponding parts of t

Page 23

3100 – 1000 µL 500 – 5000 µL 1 – 10 mL – 1000 µL1346781053421156 75a2453141112Research – variable / fix1_Research_Titel.f

Page 24

90Order numberInternationalOrder numberNorth America12 Ejection sleeve10 – 100 µL 3110 861.000 022476491200 – 1000 µL 3110 865.006 02247652113 Filling

Page 25

91Order numberInternationalOrder numberNorth America5 Cylinder attachment500 –5000 µL 3110 836.006 0224764241 – 10 mL 3111 602.009 022

Page 26

92Pipettes/spare partsOrder numberInternationalOrder numberNorth America8-channel:0.5 – 10 µL Light gray control button 3114 000.115 02245200

Page 27

93epT.I.P.S. (The packaging units stated represent the minimum ordering quantity).Order numberInternationalOrder numberNorth AmericaSilicone grease (G

Page 28

941 – 10 mL 165 mm turquoise 0030 000.765 022492098 standard (200 tips)1 – 10 mL 243 mm turquoise 0030 000.781 022492101 long (2

Page 29 - North America

95Racksplus 10 x 96 = 960 tips0.1 – 10 µL 34 mmdark gray* 0224919030.1 – 20 µL 40 mmmedium gray** 0224919110.5 – 20 µL L 46 m

Page 30

96ep Dualfilter T.I.P.S.PCR-clean, sterilein racks10 x 96 = 960 tips0.1 – 10 µL S 34 mm dark gray 0030 077.504 0224912020.1 – 10 µL M 40 mm

Page 31

97 LengthColorcodeOrder numberInternationalOrder numberNorth AmericaepT.I.P.S. LoRetention Dualfilter, PCR clean,sterile and pyrogen free, in racks,10

Page 32

98Research – Part B – Ordering information7_ResearchTeil2_E.fm Seite 98 Donnerstag, 17. April 2008 12:34 12

Page 33

117Eppendorf OfficesEppendorf South Pacific Pty. Ltd.Tel. +61 2 98 89 50 00Fax +61 2 98 89 51 11E-Mail: [email protected]: www.eppendorf.com

Page 34 - (24 tips)

4Instruction Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Bedienungsanleitung . . . . . . .

Page 35

118Eppendorf OfficesEppendorf s.r.l.Tel. +390 2 55 404 1Fax +390 2 58 013 438E-Mail: [email protected]: www.eppendorf.itEppendorf Co. Ltd.

Page 36

119Multi-channel / MehrkanalBACBACBACFig./Abb. 7Fig./Abb. 8Fig./Abb. 11Fig./Abb. 9 Press pieceNiederhalterPresseurPezzo prementeSujetadorSpringFed

Page 37 - Length

Your local distributor: www.eppendorf.com/worldwide Eppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 538 01-0 · Fax: +49 40 538 01-556E-mail: epp

Page 38

501 Avvertenze di sicurezza generali . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 511.1 Uso conforme . . . . . . . . . . . . . . . . . .

Page 39 - Eppendorf Offices

51Research – Parte A – Capitolo 11.1 Uso conformeLa Research è uno strumento di laboratorio che, in combinazione con i corrispondenti puntali per pip

Page 40

522.1 Research a volume fisso2.2 Research a volume variabileModello /VolumeAmpiezzadel passoepT.I.P.S. ErroresistematicoErrorerandomµL µL %µL%µL10202

Page 41

53 2.3 Research multicanale a volume variabile ( da 8 a 12 canali) I dati tecnici qui riportati valgono soltanto se si utilizzano puntali per pipett

Page 42 - In touch with life

54Inoltre, la parte inferiore della pipetta è girevole, consentendo così all'operatore di regolare il volume in modo individuale a seconda dell&