6 Technical data
35
Eppendorf Research® plus — Operating manual
6 Technical data
6.1 Research plus single-channel variable
Research plus variable single-channel
Model Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic Random
± % ± μL± % ± μL
0.1 - 2.5 L
Increment:
0.002 L
dark gray
0.1 - 10 L
34 mm
0.1 L ± 48 ± 0.048 ± 12 ± 0.012
0.25 L ± 12 ± 0.03 ± 6 ± 0.015
1.25 L ± 2.5 ± 0.031 ± 1.5 ± 0.019
2.5 L ± 1.4 ± 0.035 ± 0.7 ± 0.018
0.5 - 10 L
Increment:
0.01 L
medium gray
0.1 - 20 L
40 mm
0.5 L ± 8 ± 0.04 ± 5 ± 0.025
1 L ± 2.5 ± 0.025 ± 1.8 ± 0.018
5 L ± 1.5 ± 0.075 ± 0.8 ± 0.04
10 L ± 1.0 ± 0.1 ± 0.4 ± 0.04
2 - 20 L
Increment:
0.02 L
light gray
0.5 - 20 L L
46 mm
2 L ± 5 ± 0.1 ± 1.5 ± 0.03
10 L ± 1.2 ± 0.12 ± 0.6 ± 0.06
20 L ± 1.0 ± 0.2 ± 0.3 ± 0.06
2 - 20 L
Increment:
0.02 L
yellow
2 - 200 L
53 mm
2 L ± 5 ± 0.1 ± 1.5 ± 0.03
10 L ± 1.2 ± 0.12 ± 0.6 ± 0.06
20 L ± 1.0 ± 0.2 ± 0.3 ± 0.06
10 - 100 L
Increment:
0.1 L
yellow
2 - 200 L
53 mm
10 L ± 3 ± 0.3 ± 1 ± 0.1
50 L ± 1 ± 0.5 ± 0.3 ± 0.15
100 L ± 0.8 ± 0.8 ± 0.2 ± 0.2
20 - 200 L
Increment:
0.2 L
yellow
2 - 200 L
53 mm
20 L ± 2.5 ± 0.5 ± 0.7 ± 0.14
100 L ± 1 ± 1 ± 0.3 ± 0.3
200 L ± 0.6 ± 1.2 ± 0.2 ± 0.4
6
Technical data
Comments to this Manuals