4 Troubleshooting
Eppendorf Research® plus — Operating manual
12
4 Troubleshooting
4.1 Error search
Symptom Possible cause Solution
Liquid is dripping from
the tip and/or the
dispensed volume is
incorrect.
• The tip is loose or the
pipette tip is poorly fitted.
Press the tip on firmly, use
epT.I.P.S. If using 5 mL and
10 mL epDualfilter T.I.P.S., do
not use protection filters in the
pipette.
• Liquid with high vapor
pressure and/or different
density.
Wet the tip several times and
adjust the pipette for the liquid
used.
• Pipetted too quickly. Move the control button slowly.
• The tip is withdrawn from
the liquid too quickly.
Slowly remove the tip with a
time delay (approx. 3 seconds)
from the liquid.
• Liquid aspirated with
blow-out and dispensed
with blow-out.
Repeat dispensing correctly.
• The piston is soiled or
damaged.
Clean the piston, relubricate
slightly and/or replace.
• The tip cone is damaged. Replace the lower part or
channel.
• The O-rings of the tip
cones are damaged.
Replace the O-rings (only
100 L, 300 L multi-channel).
The control button jams
and does not move
smoothly.
• The piston is soiled.
• The seal is soiled.
• The pipette is blocked.
Clean the lower part.
5 mL and 10 mL sizes: replace
the protection filter.
The adjustment seal has
been removed; the
adjustment display has
been changed.
• The pipette has been
adjusted for another
liquid.
Adjust the pipette for the liquid
used (see Adjusting pipettes on
p. 22).
No spring-loading action
of the tip cone when
taking up pipette tips.
• Spring-loading action is
blocked by a locking ring.
Remove the locking ring again.
• The use of a 5 mL or
10 mL pipette.
No remedy. The tip cone does not
respond with spring-loaded action in
combination with these sizes.
4
Troubleshooting