Eppendorf 5810 R Centrifuge User Manual Page 14

  • Download
  • Add to my manuals
  • Print
  • Page
    / 95
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 13
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
14
Blood collection tube
5810 748.003
flat 3,000 × g
3 to 15mL Ø 16mm 4,000rpm
16/64 108mm/108mm 16.8 cm
Vessel
5810 721.008
flat 3,000 × g
7 to 17mL Ø 17.5mm 4,000rpm
16/64 118mm/118mm 16.8 cm
Conical tube
5810 722.004
conical 3,100×g
15mL Ø 17.5mm 4,000rpm
12/48 119mm/121mm 17.3 cm
Conical tube
5810 723.000
conical 3,100×g
50mL Ø 31mm 4,000rpm
5/20 116mm/122mm 17.3 cm
Centriprep
5810 739.004
flat 3,100×g
50mL Ø 31mm 4,000rpm
5/20 (Do not use the
aerosol-tight cap.)/
121mm
17.3 cm
Conical tube, skirted
5810 739.004
5804737.008
flat 3,100×g
50mL Ø 31mm 4,000rp
m
5/20 (Do not use the
aerosol-tight cap.)/
119mm
17.3 cm
Bottles
5825 722.000
flat 3,100×g
180 to 250mL Ø 62mm 4,000rpm
1/4 (Do not use the
aerosol-tight cap.)/
133mm
17.3 cm
Wide-neck bottle
5810 728.002
flat 3,220×g
400mL Ø 81mm 4,000rpm
1/4 (Do not use the
aerosol-tight cap.)/
133mm
18.0 cm
Vessel Vessel Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius
Eppendorf
EN
Operating manual