Centrifuge 5804/5804 R/5810/5810 R — Operating manual
18
2.4.2 Rotor A-4-62 and A-4-62-MTP (only 5810/5810 R)
Rotor A-4-62 with 250 mL rectangular bucket
Max. g-force: 3,220×g
Rotor A-4-62 Rectangular bucket
250mL
Aerosol-tight cap Max. speed: 4,000rpm
Swing-bucket rotor
with 4×250mL
rectangular buckets
Max. load per bucket
(adapter, tube and
contents):
620g
Vessel Vessel Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius
Vessel
5810 751.004
flat 3,000 × g
1.5/2mL Ø 11mm 4,000rpm
16/64 43mm/43mm 17.1 cm
Vessels
5810 750.008
flat 3,050×g
1.2 to 5mL Ø 11mm 4,000rpm
25/100 115mm/123mm 17.3 cm
Vessels
5810 752.000
flat 3,050×g
2.6 to 7mL Ø 13mm 4,000rpm
15/60 118mm/121mm 17.3 cm
Vessels
5810 753.007
flat 3,050×g
3 to 15mL Ø 16
mm 4,000rpm
12/48 116mm/121mm 17.3 cm
Vessels
5810 754.003
flat 3,050×g
7 to 17mL Ø 17.5mm 4,000rpm
12/48 114mm/118mm 17.3 cm
Conical tube
5810 755.000
conical 3,150×g
15mL Ø 17.5mm 4,000rpm
9/36 121mm/127mm 17.8 cm