Centrifuge 5804/5804 R/5810/5810 R — Operating manual
2.4 Rotors
You can operate the 5804/5804R/5810/5810R centrifuge with the following rotors. Before use of
sample tubes, please note the manufacturer's specifications with regard to centrifugation
resistance (max. rcf).
2.4.1 Rotor A-4-81 (only 5810/5810 R)
Rotor A-4-81 with 500 mL rectangular bucket
Hint!
Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.
Only use rotors which bear the name of the centrifuge (e.g. 5804R).
Max. g-force: 3,220×g
Rotor A-4-81 Rectangular bucket
500mL
Aerosol-tight cap Max. speed: 4,000rpm
Swing-bucket rotor
with 4×500mL
rectangular buckets
Max. load per bucket
(adapter, tube and
contents):
780g
Vessel Vessel Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius
Vessel
5810 745.004
flat 2,950×g
1.5/2mL Ø 11mm 4,000rpm
20/80 43mm/43mm 16.5 cm
Blood collection tube
5810 746.000
flat 3,000 × g
1.2 to 5mL Ø 11mm 4,000rpm
20/80 108mm/108mm 16.8 cm
Vessel
5810 720.001
flat 3,000 × g
2.6 to 5mL Ø 13mm 4,000rpm
25/100 107mm/108mm 16.8 cm
Vessel
5810 747.007
flat 3,000 × g
2.6 to 7mL Ø 13mm 4,000rp
m
18/72 108mm/108
mm 16.8 cm