Eppendorf 5810 R Centrifuge User Manual Page 26

  • Download
  • Add to my manuals
  • Print
  • Page
    / 95
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 25
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
26
2.4.5 Rotor A-2-DWP
Hint!
Check the load if using two fully loaded DWP plates.
Max. g-force: 2,250×g
Rotor A-2-DWP Deepwell plate
carrier
Max. speed: 3,700rpm
Swing-bucket rotor with 2 Deepwell plate
carriers
Max. load per bucket
(adapter, plate and
contents):
380g
Plate Plate Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Max. speed
Plates/slides per
adapter/rotor
Max. loading height Centrifugation radius
Micro test plate SBS adapter*
5825 718.003
flat 2,250×g
96/384 wells 4,000rpm
4/8 89mm 14.7 cm
Cell culture plate SBS adapter*
5825 718.003
flat 2,250×g
4,000rpm
4/8 89mm 14.7 cm
Deepwell plate SBS adapter*
5825 718.003
flat 2,250×g
96wells 4,000rpm
2/4 89mm 14.7 cm
Kit SBS adapter*
5825 718.003
flat 2,250×g
4,000rpm
1/2 89mm 14.7 cm
Tube in IsoRack
5825 708.008
flat 2,050×g
24 × 0.5mL Ø 6mm 3,700rpm
1/2 89mm 13.8 cm
T
ube in IsoRack
5825 709.004
flat 1,990
×g
24 × 1.5/2mL Ø 11mm 3,700rpm
1/2 89mm 13.3 cm
EN
Operating manual
Page view 25