2.4 Rotors
2.4.1 Rotor FA-45-12-17
Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.
Only use rotors which are marked with the name of the centrifuge (e.g., 5427 R).
Rotor FA-45-12-17
Aerosol-tight fixed-angle rotor
for 12 reaction tubes
Max. g-force: 20 598 × g
Max. speed: 14 000 rpm
Max. load (adapter, tube
and contents):
12 × 9.5 g
Vessels Vessel Adapters Adapter base Max. g-force:
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Centrifugation radius
Vessel 20 598 × g
5 mL Ø 17 mm 14 000 rpm
-/12 9.4 cm
Vessel
5820 768.002
open 17 530 × g
1.5 mL/2.0 mL Ø 11 mm 14 000 rpm
1/12 8.0 cm
HPLC vessel
5820 770.007
flat 16 215 × g
Ø 11 mm 14 000 rpm
1/12 7.4 cm
Cryo tube
5820 769.009
flat 18 188 × g
1.0 mL – 2.0 mL Ø 13 mm 14 000 rpm
1/12 8.3 cm