Eppendorf Centrifuge 5427 R User Manual Page 22

  • Download
  • Add to my manuals
  • Print
  • Page
    / 64
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 21
Installation
Centrifuge 5427 R
English (EN)
22
4.3 Installing the instrument
Prerequisites
The device is on a suitable lab bench.
Perform the following steps in the sequence described.
1. Let the device warm up to ambient temperature.
2. Check that the mains voltage and frequency match the requirements on the device type plate.
3. Connect the centrifuge to the mains and switch it on using the mains/power switch.
Display is active.
Lid opens automatically
4. Remove the transport securing device of the lid latch.
5. Remove the transport securing device of the motor shaft.
6. Use the details included in the scope of delivery to check that the delivery is complete.
7. Check all parts for any transport damage. Contact your dealer if any damage is found.
8. Insert the condensation water tray into the holder provided (see Fig. 2-1 on p. 7).
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a
warmer environment.
After installing the device, wait for at least 4 h. Only then connect the device to the mains
power supply.
NOTICE! Centrifuge 5427 R: compressor damage after improper transport.
Only switch on the centrifuge 4 h after installation.
Retain the packaging material and the transport protection device for subsequent transport or
storage. See also the instructions relating to transport (see p. 45).
Page view 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 63 64

Comments to this Manuals

No comments