Abb. 5: Loading the adapter 16mm75 - 100mm with tubes >119mm in length.
Fig. 5: Loading the adapter 16mm75 - 100mm with tubes >119mm in length.
Abb. 6: Symmetrical loading of the plates.
Fig. 6: Symmetrical loading of the plates.
The plate arrangement shown on the right-hand side is incorrect, as the buckets will not
swing properly.
The same principle also applies to the loading of rotor A-4-81-MTP/Flex with 4 plates.
The plates have some play in the buckets.
Abb. 7: Mixed loading of rotors
Fig. 7: Mixed loading of rotors
8. Check the loading of the bucket.
Rotor Mixing equipment
S-4-104
• 2 plate buckets (open bucket or plate
bucket)
• 2 round buckets
A-4-81/A-4-81-MTP/Flex
• 2 plate buckets (MTP or DWP bucket)
• 2 buckets for conical tubes
• 2 rectangular buckets
A-4-44
• 2 rectangular buckets
• 2 buckets for conical tubes
NOTICE!
Rotor damage due to mixed loading.
If you load the rotors A-4-62 and A-4-62-MTP with a mixed equipment, the rotors are damaged
during centrifugation.
Load all positions of the rotors A-4-62 and A-4-62-MTP with the same buckets.
Always load all 4 positions of the swing-bucket rotors.