Eppendorf 5810 R Centrifuge User Manual Page 33

  • Download
  • Add to my manuals
  • Print
  • Page
    / 95
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 32
33
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
2.4.11 Rotor S-4-104 (only 5810/5810 R)
Max. g-force: 3,214×g
Rotor S-4-104 Round bucket
750mL
Aerosol-tight cap Max. speed: 3,900rpm
Swing-bucket rotor
with 4×750mL
round buckets
Max. load per bucket
(adapter, tube and
contents):
1,000g
Vessel Vessel Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius
Vessel
5825 740.009
continuous 3,197×g
1.5/2mL Ø 11mm 3,900rpm
62/248 39mm 18.8 cm
Vessel
5825 739.000
continuous 3,214×g
5mL Ø 17mm 3,900rpm
14/56 60mm 18.9 cm
Vessel
5825 738.004
round 3,044×g
4 to 8mL Ø 13mm× 100mm 3,900rpm
23/92 108mm/115mm 17.9 cm
Vessel
5825 736.001
r
ound 3,061×
g
7,5 to 12mL Ø 16mm × 98mm 3,900rpm
20/80 114mm/119mm 18 cm
Vessel
5825 736.001
round 3,061×g
8 to 16mL Ø 16 mm 3,900rpm
7/28
Load inner bores
only (Fig. 5 on p. 54).
(Do not use the
aerosol-tight cap.)/
125mm
18 cm
Vessel
5825 743.008
round 3,044×g
9mL Ø 17.5mm×
100mm
3,900rpm
20/80 106mm/111mm 17.9 cm
Conical tube
5825 734.009
conical 3,197×g
15mL Ø 17mm× 104mm 3,900rpm
14/56 120mm/125mm 18.8 cm
EN
Operating manual
Page view 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 94 95

Comments to this Manuals

No comments