Centrifuge 5804/5804 R/5810/5810 R — Operating manual
24
2.4.4 Rotor A-2-DWP-AT (only 5810/5810 R)
Vessel
5804 760.000
flat 4,200×g
50 to 75mL Ø 35mm 5,000rpm
1/4 108mm/119mm 15.0 cm
Vessel
5804 761.006
flat 4,200×g
80 to 100mL Ø 45mm 5,000rpm
1/4 100mm/114mm 15.0 cm
Max. g-force: 3,486×g
Rotor A-2-DWP Bucket Aerosol-tight cap Max. speed: 4,500rpm
Swing-bucket rotor
with 2 aerosol-tight
buckets
(always use with
plate carrier)
Max. load per bucket
(adapter, plate and
contents):
500g
Plate Plate Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Max. speed
Plate/ slide per
adapter/rotor
Max. loading height Centrifugation radius
Micro test plate - 3,486×g
96/384 wells 4,500rpm
4/16 60mm 154mm
Cell culture plate - 3,486×g
4,500rpm
2/8 60mm 154mm
Deepwell plate flat 3,486×g
96mL
4,500rpm
1/4 67
mm 154mm
Vessel Vessel Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius