Eppendorf 5810 R Centrifuge User Manual Page 22

  • Download
  • Add to my manuals
  • Print
  • Page
    / 95
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 21
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
22
2.4.3 Rotor A-4-44
Max. g-force: 4,400×g
Rotor A-4-44 Rectangular bucket
100mL
Aerosol-tight cap Max. speed: 5,000rpm
Swing-bucket rotor
with 4×100mL
rectangular buckets
Max. load per bucket
(adapter, tube and
contents):
310g
Vessel Vessel Adapters Adapter bottom
shape
Max. g-force
Capacity Order no.
(international)
Tube diameter Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius
Vessel
5804 751.000
flat 4,100×g
1.5/2mL Ø 11mm 5,000rpm
12/48 43mm/43mm 14.8 cm
Vessels
5804 750.004
flat 4,200×g
1.2 to 5mL Ø 11mm 5,000rpm
14/56 102mm/105mm 15.0 cm
Vessels
5804 752.007
flat 4,200×g
2.6 to 7mL Ø 13mm 5,000rpm
9/36 106mm/108mm 15.0 cm
Vessels
5804 753.003
flat
4,200×
g
3 to 15mL Ø 16mm 5,000rpm
7/28 106mm/108mm 15.0 cm
Vessels
5804 754.000
flat 4,200×g
7 to 17mL Ø 17.5mm 5,000rpm
6/24 106mm/110mm 15.0 cm
Conical tube
5804 755.006
conical 4,300×g
15mL Ø 17.5mm 5,000rpm
4/16 (Do not use the
aerosol-tight cap.)/
121mm
15.5 cm
EN
Operating manual
Page view 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 94 95

Comments to this Manuals