Eppendorf 5810 R Centrifuge User Manual Page 16

  • Download
  • Add to my manuals
  • Print
  • Page
    / 95
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 15
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
16
Rotor A-4-81 with MTP/Flex carrier
Max. g-force: 2,900×g
Rotor A-4-81 MTP/Flex buckets Max. speed: 4,000rpm
Swing-bucket rotor with 4 MTP/Flex carriers Max. load per bucket
(adapter, plate and
contents):
380g
Vessel Plate Adapters Bottom shape Max. g-force
Capacity Order no.
(international)
Tube diameter Max. speed
Plates, rack or
glass slides per
adapter/rotor
Max. loading height Centrifugation radius
Micro test plate - flat 2,900×g
96/384 wells - 4,000rpm
4/16 60mm 16.3 cm
Deepwell plate - flat 2,900×g
96wells - 4,000rpm
1/4 60mm 16.3 cm
Cell culture plate - flat 2,900×g
- 4,000rpm
2/8 60mm 16.3 cm
Kit - flat 2,900×g
- 4,000rpm
1/4 60mm 16.3 cm
Tube in IsoRack
5825 708.008
flat 2,700×g
24 × 0.5mL Ø 6mm 4,000rpm
1/4 60mm 15.0 cm
Tube in IsoRack
5825 709.004
flat 2,600×g
24 × 1.5/2mL Ø 11mm 4,000rp
m
1/4 60
mm 14.6 cm
384-well PCR plate
5825 713.001
flat 2,700×g
- 4,000rpm
1/4 60mm 15.8 cm
96-well PCR plate
5825 711.009
flat 2,600×g
- 4,000rpm
1/4 60mm 16.1 cm
EN