Page 1 - Centrifuge 5430/5430 R

Register your instrument! www.eppendorf.com/myeppendorfCentrifuge 5430/5430 ROperating manual Istruzioni per l'uso

Page 2

Product descriptionCentrifuge 5430/5430 REnglish (EN)102.2 Delivery package2.2.1 Centrifuge 54302.2.2 Centrifuge 5430 R1 Emergency lid releaseOn both

Page 3 - Table of contents

UsoCentrifuge 5430/5430 RItaliano (IT)1005.7.1.2 Allentamento del coperchio rotoreUtilizzare il dispositivo di rimozione per allentare una vite del co

Page 4

UsoCentrifuge 5430/5430 RItaliano (IT)1015.7.3 QuickLock I rotori a tenuta di aerosol FA-45-48-11, FA-45-30-11, FA-45-24-11-Kit, FA-45-16-17 e S-24-11

Page 5

ManutenzioneCentrifuge 5430/5430 RItaliano (IT)1026 Manutenzione6.1 Preparazione per la pulizia/disinfezioneLa procedura descritta nel capito successi

Page 6 - English (EN)

ManutenzioneCentrifuge 5430/5430 RItaliano (IT)1036.2.1 Pulizia e disinfezione dell'apparecchio1. Aprire il coperchio. Spegnere l'apparecchi

Page 7 - 1.3 Symbols used

ManutenzioneCentrifuge 5430/5430 RItaliano (IT)1044. Pulire e disinfettare i coperchi del rotore. SOLO QuickLock: rimuovere assolutamente l'anel

Page 8 - 1.4 Abbreviations used

ManutenzioneCentrifuge 5430/5430 RItaliano (IT)1056.4 Rottura di parti in vetroIn caso di utilizzo di provette in vetro, può capitare che nella camera

Page 9 - 2.1 Main illustration

ManutenzioneCentrifuge 5430/5430 RItaliano (IT)106AVVERTENZA! Pericolo per la salute dovuto a contaminazione dell'apparecchio.1. Osservare le avv

Page 10 - 2.2 Delivery package

Risoluzione dei problemiCentrifuge 5430/5430 RItaliano (IT)1077 Risoluzione dei problemiSe con le misure proposte non è possibile eliminare l'err

Page 11 - 2.3 Features

Risoluzione dei problemiCentrifuge 5430/5430 RItaliano (IT)1087.2 Messaggi di anomaliaSe vengono visualizzati i messaggi di anomalia indicati di segui

Page 12 - 2.4 Rotors

Risoluzione dei problemiCentrifuge 5430/5430 RItaliano (IT)109Error 8Controllo della velocità• Anomalia al sistema di funzionamento• Rotore allentato•

Page 13

Product descriptionCentrifuge 5430/5430 REnglish (EN)112.3 FeaturesThe multifaceted Centrifuge 5430 has a capacity of 48 × 2.0 mL and reaches max. 30,

Page 14

Risoluzione dei problemiCentrifuge 5430/5430 RItaliano (IT)1107.3 Sblocco di emergenzaSe non è possibile aprire il coperchio della centrifuga, azionar

Page 15

Trasporto, immagazzinamento e smaltimentoCentrifuge 5430/5430 RItaliano (IT)1118 Trasporto, immagazzinamento e smaltimento8.1 Trasporto Quando si tra

Page 16

Trasporto, immagazzinamento e smaltimentoCentrifuge 5430/5430 RItaliano (IT)112

Page 17

Declarations and CertificatesCentrifuge 5430/5430 R113Centrifuge 5430/5430 Rseeon p.Fig.Tab .p.Operating manualDeclarations and Certificates

Page 24

Product descriptionCentrifuge 5430/5430 REnglish (EN)122.4 RotorsThe Centrifuge 5430/5430 R can be operated with the following rotors. Before using tu

Page 32 - 5.4.4 Manual rotor detection

Declarations and CertificatesCentrifuge 5430/5430 R127

Page 33 - 5.4.5.1 Fixed-angle rotors

Declarations and CertificatesCentrifuge 5430/5430 R128

Page 34 - 5.4.5.2 Swing-bucket rotor

AnnexCentrifuge 5430/5430 R129AnnexMenu structure of Centrifuge 5430/5430 REnglish Deutsch Français Español DisplayPrograms• Load program•Save program

Page 35

Product descriptionCentrifuge 5430/5430 REnglish (EN)13RotorFA-45-24-11-KitWith aerosol-tight QuickLock rotor lid24 spin columns or 1.5/2.0 mL tubes.W

Page 36 - 5.5 Cooling (only 5430 R)

AnnexCentrifuge 5430/5430 R130Rotor F-35-6-30: Adapters for round-bottom tubes and blood taking systems1) One tube per adapter, 6 per rotor. For the d

Page 38 - 5.5.6 Continuous cooling

Evaluate your manualGive us your feedback.www.eppendorf.com/manualfeedbackYour local distributor: www.eppendorf.com/contactEppendorf AG · 22331 Hambur

Page 39 - 5.6 Centrifuging

Product descriptionCentrifuge 5430/5430 REnglish (EN)14RotorFA-45-24-11-HS24 tubes 1.5/2.0 mL.With adapters:• 0.2 mL PCR tubes• 0.4 mL tubes• 0.5 mL t

Page 40

Product descriptionCentrifuge 5430/5430 REnglish (EN)15(1) Maximum load per rotor bore for adapter + tube + contents.(2) According to DIN 58 970 (devi

Page 41 - 5.7 Standby mode

Product descriptionCentrifuge 5430/5430 REnglish (EN)162.4.1 rcf display and calculationUse the rpm/rcf key to switch the speed of centrifugation disp

Page 42 - 5.8.1.1 Inserting the rotor

Product descriptionCentrifuge 5430/5430 REnglish (EN)17*) Centrifugation of round-bottom tubes and blood taking systems can be done using additional a

Page 43

SafetyCentrifuge 5430/5430 REnglish (EN)183Safety3.1 Intended useThe Centrifuge 5430/5430 R is intended exclusively for indoor use and for separating

Page 44 - 5.8.4 QuickLock

SafetyCentrifuge 5430/5430 REnglish (EN)19The rotors listed below, and the corresponding buckets and rotor lids, have a maximum service life of the nu

Page 45 - 6.1 Device menu

5427 901.124-04/072013Copyright© 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyri

Page 46

SafetyCentrifuge 5430/5430 REnglish (EN)203.4 Information on product liabilityIn the following cases, the designated protection of the device may be c

Page 47 - 6.2.2.2 Save program

SafetyCentrifuge 5430/5430 REnglish (EN)213.5.2 Incorrect handling of the centrifugeWARNING! Risk of injury when opening or closing the centrifuge lid

Page 48 - 6.2.3 Other menu items

SafetyCentrifuge 5430/5430 REnglish (EN)223.5.3 Incorrect handling of the rotorsWARNING! Risk of injury from improperly attached rotors and rotor lids

Page 49 - 6.2.4 Settings

SafetyCentrifuge 5430/5430 REnglish (EN)233.5.4 Extreme strain on the centrifuging tubesCAUTION! Risk of injury from overloaded tubes. Note the loadi

Page 50 - 7.2 Cleaning/disinfection

SafetyCentrifuge 5430/5430 REnglish (EN)243.5.5 Aerosol-tight centrifugation3.6 Safety notes on the deviceWARNING! Risk to health due to limited aeros

Page 51

InstallationCentrifuge 5430/5430 REnglish (EN)254 Installation4.1 Selecting the locationSelect the location for the device according to the following

Page 52

InstallationCentrifuge 5430/5430 REnglish (EN)264.3 Installing the instrumentPrerequisitesThe device is on a suitable lab bench.Perform the following

Page 53 - 7.5 Fuses

InstallationCentrifuge 5430/5430 REnglish (EN)277. Use the details included in the scope of delivery to check that the delivery is complete.8. Check a

Page 54 - 8.1 General errors

OperationCentrifuge 5430/5430 REnglish (EN)285 Operation5.1 Overview of operating controlsThe Centrifuge 5430/5430 R is available in two versions: key

Page 55 - 8.2 Error messages

OperationCentrifuge 5430/5430 REnglish (EN)29Abb. 5-2: Display of the Centrifuge 543 0/5430 RFig. 5-2: Display of the Centrifuge 5430/5430 RPlease al

Page 56

Table of contentsCentrifuge 5430/5430 REnglish (EN)3Operat ing manu alCentrifuge 5430/5430 Rseeon p.Fig.Tab .p.English (EN)Operating manualTable of co

Page 57 - 8.3 Emergency release

OperationCentrifuge 5430/5430 REnglish (EN)305.3 Configure centrifuge5.3.1 Set menu languageProceed as follows to set menu language.5.3.2 Setting the

Page 58 - 9.3 Disposal

OperationCentrifuge 5430/5430 REnglish (EN)312. Select Settings.3. Confirm your selection.4. Select Date/time.5. Confirm your selection.6. Set date.7.

Page 59 - 10.2 Ambient conditions

OperationCentrifuge 5430/5430 REnglish (EN)325.4 Preparing for centrifugation5.4.1 Switching on the centrifuge1. Switch on the centrifuge using the ma

Page 60 - 10.4 Application parameters

OperationCentrifuge 5430/5430 REnglish (EN)335.4.5 Loading the rotor5.4.5.1 Fixed-angle rotorsTo load the rotor, proceed as follows:1. Check the maxim

Page 61

OperationCentrifuge 5430/5430 REnglish (EN)345.4.5.2 Swing-bucket rotorPrerequisites• A combination of rotor, carrier and adapter, approved by Eppendo

Page 62 - 11.2 Centrifuge 5430 R

OperationCentrifuge 5430/5430 REnglish (EN)356. Load the buckets symmetrically.Abb. 5-3: Incomplete, but symmetric loading of the buckets. The pegs of

Page 63

OperationCentrifuge 5430/5430 REnglish (EN)365.5 Cooling (only 5430 R)5.5.1 Temperature adjustment Set the temperature using the temp arrow keys betw

Page 64

OperationCentrifuge 5430/5430 REnglish (EN)375.5.5 FastTemp proYou can have the previously described temperature control run FastTemp (see p. 36) star

Page 65

OperationCentrifuge 5430/5430 REnglish (EN)385.5.6 Continuous coolingIf the rotor stops, the rotor chamber will be maintained at the target temperatur

Page 66

OperationCentrifuge 5430/5430 REnglish (EN)395.6 CentrifugingEach of the centrifuging variants described here must be preceded by the preparation desc

Page 67 - 11.4 Accessories

Table of contentsCentrifuge 5430/5430 REnglish (EN)45.5 Cooling (only 5430 R) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 68 - 11.4.2 Other accessories

OperationCentrifuge 5430/5430 REnglish (EN)405.6.2 Centrifuging in continuous operationPerform the following steps in the sequence described.1. Use th

Page 69

OperationCentrifuge 5430/5430 REnglish (EN)415.6.4 Removing the rotor1. Turn the rotor nut counterclockwise using the supplied rotor key.Rotor FA-45-2

Page 70

OperationCentrifuge 5430/5430 REnglish (EN)425.8 User instructions on rotors5.8.1 Rotor F-35-6-30: rotor removal tool5.8.1.1 Inserting the rotor Prere

Page 71 - 2.1 Illustrazione generale

OperationCentrifuge 5430/5430 REnglish (EN)435.8.2 Rotor A-2-MTP5.8.2.1 Inserting the rotor 5.8.2.2 Position and remove the upper shell of the wind sh

Page 72

OperationCentrifuge 5430/5430 REnglish (EN)445.8.4 QuickLock Aerosol-tight rotors FA-45-48-11, FA-45-30-11, FA-45-24-11-Kit, FA-45-16-17 and S-24-11-A

Page 73 - 2.2 Rotori

Operating controls and functionCentrifuge 5430/5430 REnglish (EN)456 Operating controls and function6.1 Device menuMost menu levels contain the additi

Page 74

Operating controls and functionCentrifuge 5430/5430 REnglish (EN)465430: only for Centrifuge 54305430 R: only for Centrifuge 5430 R •Volume–Cancel–Sav

Page 75

Operating controls and functionCentrifuge 5430/5430 REnglish (EN)476.2 Settings in the device menu6.2.1 ProgramsThe Centrifuge 5430/5430 R has more th

Page 76

Operating controls and functionCentrifuge 5430/5430 REnglish (EN)486.2.3 Other menu itemsMenu item/meaning Setting Function DisplaySoft rampReduce spe

Page 77

Operating controls and functionCentrifuge 5430/5430 REnglish (EN)496.2.4 SettingsMenu item/meaning Setting Function DisplayDisplaySelect standard disp

Page 78 - 3.3 Limiti di applicazione

Table of contentsCentrifuge 5430/5430 REnglish (EN)511 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 79

MaintenanceCentrifuge 5430/5430 REnglish (EN)507 Maintenance7.1 Prepare cleaning/disinfection Clean all accessible surfaces of the device and the acc

Page 80

MaintenanceCentrifuge 5430/5430 REnglish (EN)517.2.1 Cleaning and disinfecting the device1. Open the lid. Switch off the device with the mains/power s

Page 81

MaintenanceCentrifuge 5430/5430 REnglish (EN)524. Clean and disinfect the rotor lids. ONLY QuickLock: It is imperative that the sealing ring be remov

Page 82 - 3.5.3 Uso errato dei rotori

MaintenanceCentrifuge 5430/5430 REnglish (EN)537.4 Glass breakageWhen using glass tubes there is a risk of glass breakage in the rotor chamber. The re

Page 83

TroubleshootingCentrifuge 5430/5430 REnglish (EN)548 TroubleshootingIf you cannot remedy an error with the recommended measures, please contact your l

Page 84

TroubleshootingCentrifuge 5430/5430 REnglish (EN)558.2 Error messagesIf one of the following error messages appears, proceed as follows:1. Remove faul

Page 85

TroubleshootingCentrifuge 5430/5430 REnglish (EN)56Error 16 to Error 17Electronics faultElectronics fault.  Switch centrifuge off and back on again a

Page 86 - 4 Installazione

TroubleshootingCentrifuge 5430/5430 REnglish (EN)578.3 Emergency releaseIf the centrifuge lid cannot be opened, you can activate the emergency release

Page 87

Transport, storage and disposalCentrifuge 5430/5430 REnglish (EN)589 Transport, storage and disposal9.1 Transport Remove the rotor from the centrifug

Page 88

Technical dataCentrifuge 5430/5430 REnglish (EN)5910 Technical data10.1 Power supplyCentrifuge 5430Centrifuge 5430 R10.2 Ambient conditionsMains conne

Page 89

Table of contentsCentrifuge 5430/5430 REnglish (EN)6

Page 90

Technical dataCentrifuge 5430/5430 REnglish (EN)6010.3 Weight/dimensionsCentrifuge 5430Centrifuge 5430 R*) The noise level was measured frontally in a

Page 91 - 5.3.2 Inserimento del rotore

Technical dataCentrifuge 5430/5430 REnglish (EN)61Acceleration times and braking times of Centrifuge 5430/5430 RRotor Lowest achievable temperature -1

Page 92 - 5.3.5 Caricamento del rotore

Ordering InformationCentrifuge 5430/5430 REnglish (EN)6211 Ordering Information11.1 Centrifuge 543011.2 Centrifuge 5430 ROrder no. (International)Orde

Page 93 - 5.3.5.2 Rotori basculanti

Ordering InformationCentrifuge 5430/5430 REnglish (EN)6311.3 Rotors, rotor lids and seals11.3.1 Rotors with QuickLock rotor lidRotor FA-45-48-11Rotor

Page 94

Ordering InformationCentrifuge 5430/5430 REnglish (EN)64Rotor FA-45-24-11 KitRotor FA-45-16-17Rotor S-24-11-ATOrder no. (International)Order no. (Nort

Page 95 - 5.4.4 FastTemp

Ordering InformationCentrifuge 5430/5430 REnglish (EN)6511.3.2 Rotors with rotor lid threadRotor F-45-48-11Rotor FA-45-30-11Rotor F-45-30-11Rotor FA-4

Page 96 - 5.4.5 FastTemp pro

Ordering InformationCentrifuge 5430/5430 REnglish (EN)66Rotor FA-45-24-11-KitRotor F-45-64-5-PCRRotor F-45-18-17-CryoRotor F-35-6-30 Order no. (Intern

Page 97 - 5.5 Centrifugazione

Ordering InformationCentrifuge 5430/5430 REnglish (EN)6711.3.3 Rotors with rotor lid for attachingA-2-MTP rotor11.4 Accessories11.4.1 AdapterOrder no.

Page 98 - 5.5.3 Rimozione del rotore

Ordering InformationCentrifuge 5430/5430 REnglish (EN)6811.4.2 Other accessories11.4.3 Fuses for Centrifuge 54305427 743.006 5427743006 13 × 90-110, s

Page 99 - 5.6 Modalità Standby

IndiceCentrifuge 5430/5430 RItaliano (IT)69Istruzioni per l' uso Centrifuge 5430/5430 Rvedia pag.Fig.Ta b.a pag.Italiano (IT)Istruzioni per l&apo

Page 100 - 5.7.2 Rotore A-2-MTP

Operating instructionsCentrifuge 5430/5430 REnglish (EN)71 Operating instructions1.1 Using this manual Read this operating manual completely before u

Page 101 - 5.7.3 QuickLock

IndiceCentrifuge 5430/5430 RItaliano (IT)705.7 Istruzioni per l'uso dei rotori . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 102 - 6 Manutenzione

Avvertenze per l'utilizzoCentrifuge 5430/5430 RItaliano (IT)711 Avvertenze per l'utilizzo1.1 Impiego delle presenti istruzioni Prima di met

Page 103

Descrizione del prodottoCentrifuge 5430/5430 RItaliano (IT)72Abb. 2-2: Veduta anteriore e posteriore della Centrifuga 5430 RFig. 2-2: Veduta anterior

Page 104

Descrizione del prodottoCentrifuge 5430/5430 RItaliano (IT)732.2 RotoriLa Centrifuge 5430/5430 R può essere azionata con i seguenti rotori. Prima dell

Page 105 - 6.5 Fusibili

Descrizione del prodottoCentrifuge 5430/5430 RItaliano (IT)74RotoreF-45-30-11 con coperchio rotore in polipropilene30 provette di reazione da 1,5/2,0

Page 106

Descrizione del prodottoCentrifuge 5430/5430 RItaliano (IT)75RotoreS-24-11-ATcon coperchio rotore QuickLock a tenuta di aerosol24 provette di reazione

Page 107 - 7.1 Anomalie generiche

Descrizione del prodottoCentrifuge 5430/5430 RItaliano (IT)76RotoreF-45-64-5-PCR64 provette PCR (0,2 mL)o8 strisce PCR 5/8,ciascuna con adattatore.13.

Page 108 - 7.2 Messaggi di anomalia

Descrizione del prodottoCentrifuge 5430/5430 RItaliano (IT)77(1) Carico massimo per alesaggio rotore per adattatore + recipiente + contenuto.(2) In ba

Page 109

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)783 Avvertenze di sicurezza generali3.1 Uso conformeLa Centrifuge 5430/5430 R è con

Page 110 - 7.3 Sblocco di emergenza

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)79I seguenti rotori, compresi i relativi cestelli, il supporto e il coperchio del r

Page 111 - 8.3 Smaltimento

Operating instructionsCentrifuge 5430/5430 REnglish (EN)81.4 Abbreviations usedMTPMicro test platePCRPolymerase chain reactionPTFEPolytetrafluorethyle

Page 112 - Italiano (IT)

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)803.4 Note sulla responsabilità da prodottoNei seguenti casi la responsabilità da p

Page 113 - Declarations and Certificates

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)81AVVERTENZA! Danni alla salute dovuti a liquidi infettivi e germi patogeni. Per i

Page 114

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)823.5.2 Uso errato della centrifuga3.5.3 Uso errato dei rotoriAVVISO! Danni a causa

Page 115

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)833.5.4 Sollecitazione eccessiva delle provette durante la centrifugazioneAVVISO! I

Page 116

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)843.5.5 Centrifugazione a tenuta di aerosolAVVISO! Le provette si riscaldano.Nelle

Page 117

Avvertenze di sicurezza generaliCentrifuge 5430/5430 RItaliano (IT)853.6 Avvertenze di sicurezza sull'apparecchioIllustrazione Significato Ubicaz

Page 118

InstallazioneCentrifuge 5430/5430 RItaliano (IT)864 Installazione4.1 Scelta dell'ubicazioneScegliere l'ubicazione dell'apparecchio in b

Page 119

InstallazioneCentrifuge 5430/5430 RItaliano (IT)873. Collegare la centrifuga alla rete e attivarla con l'interruttore di rete sul retro dell&apos

Page 120

UsoCentrifuge 5430/5430 RItaliano (IT)885Uso5.1 Panoramica elementi di comandoAbb. 5-1: Tastiera di comando e display de lla Centrifuge 5430/5430 R (v

Page 121

UsoCentrifuge 5430/5430 RItaliano (IT)89Abb. 5-2: Display della Centrifuge 5430/5430 RFig. 5-2: Display della Centrifuge 5430/5430 R1 Numero di progra

Page 122

Product descriptionCentrifuge 5430/5430 REnglish (EN)92 Product description2.1 Main illustrationAbb. 2-1: Front and rear view of Centrifuge 5430Fig. 2

Page 123

UsoCentrifuge 5430/5430 RItaliano (IT)905.2 Configurazione della centrifuga5.2.1 Impostazione della lingua del menuAprire il menu di selezione della l

Page 124

UsoCentrifuge 5430/5430 RItaliano (IT)915.3 Preparazione della centrifugazione5.3.1 Accensione della centrifuga1. Accendere la centrifuga agendo sull&

Page 125

UsoCentrifuge 5430/5430 RItaliano (IT)925.3.5 Caricamento del rotore5.3.5.1 Rotori ad angolo fissoPer caricare il rotore, procedere nel modo indicato

Page 126

UsoCentrifuge 5430/5430 RItaliano (IT)935.3.5.2 Rotori basculantiPremessa• Una combinazione di rotore, supporto e adattatore approvata da Eppendorf.•

Page 127

UsoCentrifuge 5430/5430 RItaliano (IT)945.3.6 Chiusura del coperchio della centrifuga1. Controllare il corretto fissaggio del rotore e del relativo co

Page 128

UsoCentrifuge 5430/5430 RItaliano (IT)955.4.4 FastTempCon questa funzione si avvia direttamente un ciclo di controllo della temperatura senza campioni

Page 129

UsoCentrifuge 5430/5430 RItaliano (IT)965.4.5 FastTemp proÈ possibile far partire in automatico a una certa ora il ciclo di controllo della temperatur

Page 130

UsoCentrifuge 5430/5430 RItaliano (IT)975.4.6 Raffreddamento permanenteIn caso di arresto del rotore, la camera rotore viene mantenuta alla temperatur

Page 131

UsoCentrifuge 5430/5430 RItaliano (IT)98Fine della centrifugazione• Una volta trascorso il tempo impostato, la centrifuga si ferma automaticamente. Du

Page 132 - Evaluate your manual

UsoCentrifuge 5430/5430 RItaliano (IT)995.6 Modalità StandbyLa centrifuga passa automaticamente dallo stato operativo alla modalità Standby quando i s