Technical specifications
Eppendorf SOP
English (EN)
9.4 Multipette
The following specifications for the Multipette plus, Multipette stream /
Multipette Xstream, Multipette M4 apply to the following conditions:
Using the Combitips advanced
Liquid: Distilled or deionized water
Reference temperature: 20 to 25 °C, ±0.5 °C
Number of determinations: 10 according to EN ISO 8655, with the original
Eppendorf Combitip advanced
Multipette stream / Multipette Xstream:
Volume test in "DIS" mode
Set speed level: 7
9.4.1 MultipettePlus
Combitip advanced Testing volume Error limits
Error
Systematic error Random error
± % ± µL ± % ± µL
0.1 mL
(beige piston)
2 µL ±1.6 ±0.032 ±3.0 ±0.06
20 µL ±1.0 ±0.2 ±2.0 ±0.4
0.2 mL
(blue piston)
4 µL ±1.3 ±0.052 ±2.0 ±0.08
40 µL ±0.8 ±0.32 ±1.5 ±0.6
0.5 mL 10 µL ±0.9 ±0.09 ±1.5 ±0.15
100 µL ±0.8 ±0.8 ±0.6 ±0.6
1 mL 20 µL ±0.9 ±0.18 ±0.9 ±0.18
200 µL ±0.6 ±1.2 ±0.4 ±0.8
2.5 mL 50 µL ±0.8 ±0.4 ±0.8 ±0.4
500 µL ±0.5 ±2.5 ±0.3 ±1.5
5 mL 100 µL ±0.6 ±0.6 ±0.6 ±0.6
1,000 µL ±0.5 ±5.0 ±0.25 ±2.5
10 mL 200 µL ±0.5 ±1.0 ±0.6 ±1.2
2 000 µL ±0.5 ±10 ±0.25 ±5.0
25 mL
(blue adapter)
500 µL ±0.4 ±2.0 ±0.6 ±3.0
5,000 µL ±0.3 ±15 ±0.25 ±12.5
50 mL
(dark gray adapter)
1,000 µL ±0.3 ±3.0 ±0.5 ±5.0
10,000 µL ±0.3 ±30 ±0.3 ±30