Eppendorf SOP
Standard Operating Procedure for Pipettes
Page view 0
1 2 3 4 5 6 ... 63 64

Summary of Contents

Page 1 - Eppendorf SOP

rating Procedure for PipettesOPEN)Operating Procedure for Pipettes Register your instrument! www.eppendorf.com/myeppendorfEppendorf SOPStandard Operat

Page 2

10CalibrationEppendorf SOPEnglish (EN)2.2.3 Dispensing the test volume into the weighing vessel1. Rest the filled tip up against the wall of the weigh

Page 3 - Table of contents

11EvaluationEppendorf SOPEnglish (EN)3 Evaluation3.1 Calculating the systematic errorMean value of the dispensed volume:n = Number of measured valuesT

Page 4

12Sterilization and cleaningEppendorf SOPEnglish (EN)4 Sterilization and cleaning4.1 SterilizationParameter for autoclaving• 121 °C• 20 minutes• 1 bar

Page 5

13Sterilization and cleaningEppendorf SOPEnglish (EN)4.1.3 Research and Research proWith the Research and Research pro pipettes, the lower section is

Page 6 - 1.1 Balances

14Sterilization and cleaningEppendorf SOPEnglish (EN)4.1.7 Xplorer and Xplorer plusThe lower part of theXplorer and Xplorer plus pipettes can be autoc

Page 7 - 1.2 Measuring station

15Leakage checkEppendorf SOPEnglish (EN)5 Leakage check5.1 Pipette leakage checkIn order to carry out a leakage check on a pipette, follow the procedu

Page 8 - 2.1 Scope of testing

16TroubleshootingEppendorf SOPEnglish (EN)Liquid cannot be aspirated correctly or dripping occurs. • Leakage in the Combitip advanced.  Replace the C

Page 9 - 2.2 Measurement

17AdjustmentEppendorf SOPEnglish (EN)7Adjustment7.1 General informationPrior to delivery, all pipettes are adjusted with distilled or de-ionized, dega

Page 10 - English (EN)

18AdjustmentEppendorf SOPEnglish (EN)7.2.1 Procedure Requirements Device, original Eppendorf tip, test liquid and ambient air must have the same tempe

Page 11 - 3 Evaluation

19AdjustmentEppendorf SOPEnglish (EN)7.2.2 BiomasterAuxiliary equipment • Supplied pipette wrench (order no. 4910 092.001) 7.2.3 Reference variableAux

Page 12 - 4.1 Sterilization

AESOP13640-08/052013Copyright© 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyrigh

Page 13 - 4.1.6 Varispenser

20AdjustmentEppendorf SOPEnglish (EN)7.2.4 Reference 2 adjustable-volume7.2.4.1 User adjustmentChanging the adjustment will change the volume by a spe

Page 14 - 4.2 Cleaning

21AdjustmentEppendorf SOPEnglish (EN)7. Attach the red adjustment seal after the inspections.If the adjustment only applies to a specific liquid, mark

Page 15 - 6 Troubleshooting

22AdjustmentEppendorf SOPEnglish (EN)1. Check whether the adjustment display is set to "0".If the adjustment display is not set to “0”, it m

Page 16 - Troubleshooting

23AdjustmentEppendorf SOPEnglish (EN)5. Turn the adjustment tool to set the volume display. Set the volume display (from the volume setting used durin

Page 17 - 7.1 General information

24AdjustmentEppendorf SOPEnglish (EN)7.2.5 Research variable Auxiliary equipment • Supplied pipette wrench (order no. 3111 501.016) • Supplied red CAL

Page 18 - 7.2.1 Procedure

25AdjustmentEppendorf SOPEnglish (EN)7.2.6 Research plusadjustable volume - adjustment to environmental parametersAdjustment change for specific liqui

Page 19 - 7.2.3 Reference variable

26AdjustmentEppendorf SOPEnglish (EN)7.2.7 Research plus adjustable volume - change to the factory adjustmentAuxiliary equipment • Supplied safety plu

Page 20 - 7.2.4.1 User adjustment

27AdjustmentEppendorf SOPEnglish (EN)2. Provide the volume setting ring and the ejector with a common mark with a pen. This mark is used for orientati

Page 21 - Adjustment

28AdjustmentEppendorf SOPEnglish (EN)4. Continue to keep the ejector pressed. Insert the safety plug tool in such a way that the counter locking mecha

Page 22

29AdjustmentEppendorf SOPEnglish (EN)6. This results in the following approximate volume changes:The values mentioned are theoretical values and are f

Page 23

3Table of contentsEppendorf SOPEnglish (EN)Table of contents1 Test and pipette conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 24 - 7.2.5 Research variable

30AdjustmentEppendorf SOPEnglish (EN)9. Document the changes made and the measurements conducted. Remove the mark on the volume setting ring and on th

Page 25

31AdjustmentEppendorf SOPEnglish (EN)7.2.8 Research Multi-channel Auxiliary equipment • Supplied pipette wrench (order no. 3111 501.016) •Supplied red

Page 26

32AdjustmentEppendorf SOPEnglish (EN)7.2.10 Varispenser plus 7.2.11 Xplorer and Xplorer plus Turn the fine adjustment toward + or –. A revolution cor

Page 27

33AdjustmentEppendorf SOPEnglish (EN)7.3 Adjusting fixed-volume pipettes The adjustment process for fixed-volume pipettes is identical to the adjustm

Page 28

34AdjustmentEppendorf SOPEnglish (EN)7.3.2 Reference2Adjustment is carried out as described for the Reference 2 adjustable-volume (see p. 20). 7.3.3 R

Page 29

35AdjustmentEppendorf SOPEnglish (EN)7.3.4 Research plus fixed Adjustment is carried out as described for the Research plus adjustable-volume(see p. 2

Page 30

36Factor Z for distilled waterEppendorf SOPEnglish (EN)8 Factor Z for distilled water8.1 Factor Z overview tableFactor Z (µl/mg) in accordance with EN

Page 31 - 7.2.8 Research Multi-channel

37Technical specificationsEppendorf SOPEnglish (EN)9 Technical specificationsThe following specifications are valid for the following conditions:9.1 F

Page 32 - 7.2.10 Varispenser plus

38Technical specificationsEppendorf SOPEnglish (EN)9.1.2 Reference 2 fixedFixed-volume Reference2 Model Test tipepT.I.P.S. Color codeVolume rangeLengt

Page 33 - 7.3.1 Reference fixed

39Technical specificationsEppendorf SOPEnglish (EN)9.1.3 Research fixedModel Test tipepT.I.P.S. Color codeVolume rangeLengthError limitsErrorSystemati

Page 34 - 7.3.3 Research fixed

4Table of contentsEppendorf SOPEnglish (EN)5 Leakage check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 35 - 7.3.4 Research plus fixed

40Technical specificationsEppendorf SOPEnglish (EN)9.1.4 Research plus fixedResearch plus fixed volumeModel Test tipepT.I.P.S. Color codeVolume rangeL

Page 36 - 8.1 Factor Z overview table

41Technical specificationsEppendorf SOPEnglish (EN)9.2 Adjusting adjustable volume pipettes 9.2.1 Reference variable Model Test tipepT.I.P.S. Color c

Page 37 - 9.1 Fixed-volume pipettes

42Technical specificationsEppendorf SOPEnglish (EN)9.2.2 Reference 2 adjustable-volume Reference2 single-channel variableModel Test tipepT.I.P.S. Colo

Page 38 - 9.1.2 Reference 2 fixed

43Technical specificationsEppendorf SOPEnglish (EN)Reference2 single-channel variableModel Test tipepT.I.P.S. Color codeVolume rangeLengthTesting volu

Page 39 - 9.1.3 Research fixed

44Technical specificationsEppendorf SOPEnglish (EN)9.2.3 Research adjustable-volume Model Test tipepT.I.P.S. Color codeVolume rangeLengthTesting volum

Page 40 - 9.1.4 Research plus fixed

45Technical specificationsEppendorf SOPEnglish (EN)9.2.4 Research proModel Test tipepT.I.P.S. Color codeVolume rangeLengthTesting volumeError limitsEr

Page 41 - 9.2.1 Reference variable

46Technical specificationsEppendorf SOPEnglish (EN)9.2.5 Research plus adjustable-volume Research plus adjustable single-channelModel Test tipepT.I.P.

Page 42 - Technical specifications

47Technical specificationsEppendorf SOPEnglish (EN)9.2.6 Biomaster9.2.7 Varipette0.5 µL – 5 mLIncrement:0.005 mLPurple0.1 µL – 5 mL120 mm0.5 mL ±2.4

Page 43

48Technical specificationsEppendorf SOPEnglish (EN)9.2.8 Xplorer and Xplorer pluspipetteIncrementPipette tipColor codeVolume rangeLengthTesting volume

Page 44

49Technical specificationsEppendorf SOPEnglish (EN)9.3 Multi-channel pipettes 9.3.1 ResearchModel Test tipepT.I.P.S. Color codeVolume rangeLengthVol

Page 45 - 9.2.4 Research pro

5Table of contentsEppendorf SOPEnglish (EN)9 Technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Page 46

50Technical specificationsEppendorf SOPEnglish (EN)9.3.2 Research proModel Test tipepT.I.P.S. Color codeVolume rangeLengthTesting volumeError limitsEr

Page 47 - 9.2.7 Varipette

51Technical specificationsEppendorf SOPEnglish (EN)9.3.3 Research plusResearch plus adjustable multi-channelModel Test tipepT.I.P.S. Color codeVolume

Page 48

52Technical specificationsEppendorf SOPEnglish (EN)9.3.4 Reference2Reference2 Variable multi-channel (8-channel/12-channel)Model Test tipepT.I.P.S. Co

Page 49 - 9.3 Multi-channel pipettes

53Technical specificationsEppendorf SOPEnglish (EN)9.3.5 Xplorer and Xplorer pluspipetteIncrementPipette tipColor codeVolume rangeLengthTesting volume

Page 50 - 9.3.2 Research pro

54Technical specificationsEppendorf SOPEnglish (EN)9.4 Multipette The following specifications for the Multipette plus, Multipette stream / Multipett

Page 51 - 9.3.3 Research plus

55Technical specificationsEppendorf SOPEnglish (EN)9.4.2 MultipetteM4Combitip advanced Testing volume Error limitsErrorSystematic error Random error±

Page 52 - 9.3.4 Reference2

56Technical specificationsEppendorf SOPEnglish (EN)9.4.3 Multipette stream / Multipette XstreamCombitip advancedVolume range Testing volumeError limit

Page 53

57Technical specificationsEppendorf SOPEnglish (EN)9.5 Varispenser / Top Buret The following specifications for Varispenser and Top Buret are vali

Page 54 - 9.4 Multipette

58Technical specificationsEppendorf SOPEnglish (EN)9.6.1 Air-cushion pipettes, fixed and adjustable volume9.6.2 Positive displacement pipettes (Biomas

Page 55 - 9.4.2 MultipetteM4

59Technical specificationsEppendorf SOPEnglish (EN)9.6.3 Dispenser (Multipette)Nominal volume Error limitsMeasurement errorsSystematic Random± % ± µL

Page 56

6Test and pipette conditionsEppendorf SOPEnglish (EN)1 Test and pipette conditionsFor the calibration or adjustment of pipettes, the scales and measur

Page 57 - 9.5.2 Top Buret

60Technical specificationsEppendorf SOPEnglish (EN)9.6.4 Single-stroke dispenser (Varispenser)9.6.5 Piston stroke buretsNominal volume Error limitsMea

Page 58

61IndexEppendorf SOPEnglish (EN)IndexAAccuracy... 6Adjustable-volume pipettesAdjust...

Page 59 - 9.6.3 Dispenser (Multipette)

62IndexEppendorf SOPEnglish (EN)RRandom error... 11SScope of testingAdjusting adjustable volume pipettes ...

Page 61

Evaluate your manualGive us your feedback.www.eppendorf.com/manualfeedbackYour local distributor: www.eppendorf.com/contactEppendorf AG · 22331 Hambur

Page 62

7Test and pipette conditionsEppendorf SOPEnglish (EN)1.2 Measuring station1.2.1 EvaporationTake evaporation protection into account while carrying out

Page 63

8CalibrationEppendorf SOPEnglish (EN)2 Calibration2.1 Scope of testingNote concerning the nominal volume:The nominal volume of an adjustable-volume pi

Page 64 - Evaluate your manual

9CalibrationEppendorf SOPEnglish (EN)2.1.4 Bottle-top dispenser and Top Buret For bottle-top dispensers and the Top Buret, check the nominal volume wi