Operation
Centrifuge 5427 R
English (EN)
32
5.5.5 Continuous cooling
Continuous cooling
Continuous cooling maintains the rotor chamber at the set temperature if the rotor stops.
During continuous cooling the display shows the set temperature.
To prevent the rotor chamber from freezing or condensation from forming, the temperature does not go
below 4 °C , irrespective of the set temperature.
If the rotor stops, temperature control is slower than during centrifugation or a temperature control run.
ECO shut-off
Continuous cooling is switched off if the centrifuge is not used for longer than the preset time. The
centrifuge switches to standby mode.
Default setting: Continuous cooling ends after 8 h.
Continuous cooling can be limited to 1 h, 2 h or 4 h.
ECO shut-off can be switched off (continuous cooling set to endless operation).
Limit continuous cooling to 1 h (2 h, 4 h).
1. Press the menu/enter key.
2. Select TEMP using the arrow key. Confirm with the menu/enter key.
3. Select 1 h (2 h, 4 h) using the arrow key. Confirm with the menu/enter key.
4. To activate ECO shut-off after 1 h, 2 h or 4 h, open and close again the centrifuge lid after the end of the
run.
The device switches to standby mode after 1 h, 2 h or 4 h. The display shows EP.
5.5.6 Endless operation of continuous cooling
The ECO shut-off function can be switched off. Continuous cooling is changed to endless operation.
Endless operation can shorten the service life of the compressor.
The rotor chamber may freeze.
1. Press the menu/enter key.
2. Select TEMP using the arrow key. Confirm with the menu/enter key.
3. Select oo using the arrow key. Confirm with the menu/enter key.
Ending continuous cooling
4. Open the centrifuge lid to end continuous cooling.
If the centrifuge lid is not opened and closed again after the run, ECO shut-off engages
after 8 h (default setting).
Open the centrifuge lid to end continuous cooling early.