Technical data
Eppendorf Reference
®
2
English (EN)
35
7.3 Reference 2 fixed
Reference 2 fixed volume
Model Test tip
epT.I.P.S. color code
Volume range
Length
Eppendorf AG error limits
Error
Systematic error Random error
± µL ± µL
1 µL Dark gray
0.1 µL – 10 µL
34 mm
±2.5 ±0.025 ±1.8 ±0.018
2 µL ±2.0 ±0.04 ±1.2 ±0.024
5 µL Medium gray
0.1 µL – 20 µL
40 mm
±1.2 ±0.06 ±0.6 ±0.03
10 µL ±1.0 ±0.1 ±0.5 ±0.05
20 µL Light gray
0.5 µL – 20 µL L
46 mm
±0.8 ±0.16 ±0.3 ±0.06
10 µL Yellow
2 µL 200 µL
53 mm
±1.2 ±0.12 ±0.6 ±0.06
20 µL ±1.0 ±0.2 ±0.3 ±0.06
25 µL ±1.0 ±0.25 ±0.3 ±0.075
50 µL ±0.7 ±0.35 ±0.3 ±0.15
100 µL ±0.6 ±0.6 ±0.2 ±0.2
200 µL ±0.6 ±1.2 ±0.2 ±0.4
200 µL Blue
50 µL – 1000 µL
71 mm
±0.6 ±1.2 ±0.2 ±0.4
250 µL ±0.6 ±1.5 ±0.2 ±0.5
500 µL ±0.6 ±3.0 ±0.2 ±1.0
1000 µL ±0.6 ±6.0 ±0.2 ±2.0
Fixed-volume Reference 2
Model Test tip
epT.I.P.S. color code
Volume range
Length
Eppendorf AG error limits
Error
Systematic error Random error
± mL ± mL
2.0 mL Red
0.5 mL – 2.5 mL
115 mm
±0.6 ±0.012 ±0.2 ±0.004
2.5 mL ±0.6 ±0.015 ±0.2 ±0.005