Eppendorf Pipette Reference 2 User Manual Page 1

Browse online or download User Manual for Tools Eppendorf Pipette Reference 2. Eppendorf Pipette Reference 2 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Eppendorf Reference

nualeference® 2EN)g manual Register your instrument! www.eppendorf.com/myeppendorfEppendorf Reference® 2Operating manual

Page 2

Product descriptionEppendorf Reference® 2English (EN)102.3.2 Reference 2 Multi-channelFor an explanation of the upper part of the pipette, refer to th

Page 3 - Table of contents

Product descriptionEppendorf Reference® 2English (EN)112.4 MaterialsThe Reference 2 components that can be accessed by the user are made of the follow

Page 4

Product descriptionEppendorf Reference® 2English (EN)122.5 Storing the pipetteThe pipette can be safely stored in a pipette carousel or a wall mountin

Page 5 - 1.3 Symbols used

SafetyEppendorf Reference® 2English (EN)133 Safety3.1 Intended use3.2 Warnings for intended useThe intended use is described in a separate document en

Page 6 - Operating instructions

SafetyEppendorf Reference® 2English (EN)14NOTICE! Damage to device from missing pipette tips. Use the Reference 2 with fitted pipette tips only.  Wh

Page 7 - 2.1 Features

OperationEppendorf Reference® 2English (EN)154 Operation4.1 Setting the volume (only Reference 2 variable)4.2 Using pipette tipsThe liquid to be dispe

Page 8 - 2.2 Delivery package

OperationEppendorf Reference® 2English (EN)164.3 Deactivating the tip cone spring loadingFor the following dispensing tasks, it may be helpful to deac

Page 9 - 2.3 Main illustration

TroubleshootingEppendorf Reference® 2English (EN)175 Troubleshooting5.1 Error searchSymptom/message Cause RemedyLiquid is dripping from the tip and/or

Page 10 - Reference 2

TroubleshootingEppendorf Reference® 2English (EN)18Regularly check the precision and accuracy of the Reference 2 to prevent dispensing errors. You can

Page 11 - 2.4 Materials

MaintenanceEppendorf Reference® 2English (EN)196 Maintenance• All single and multi-channel lower parts are wear parts. Clean them after contamination,

Page 12 - 2.5 Storing the pipette

4920 900.017-01/052013 Copyright© 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyri

Page 13 - 3.2 Warnings for intended use

MaintenanceEppendorf Reference® 2English (EN)206.2 Sterilizing or disinfecting the pipette6.2.1 AutoclavingAll Reference 2 pipettes are completely ste

Page 14

MaintenanceEppendorf Reference® 2English (EN)216.2.2 Disinfection Carefully wipe off the outer surfaces with disinfectant, DNA/RNA decontamination ag

Page 15 - 4.2 Using pipette tips

MaintenanceEppendorf Reference® 2English (EN)226.4 Disassembling and assembling the pipette lower part6.4.1 Single-channel lower part up to 1000 µL6.4

Page 16 - 4.5 Dispensing liquid

MaintenanceEppendorf Reference® 2English (EN)236.4.1.3 Opening and assembling the lower part (≤ 1000 µL) To avoid confusion of parts, only disassemble

Page 17 - 5.1 Error search

MaintenanceEppendorf Reference® 2English (EN)246.4.2 Single-channel lower part Applies to the following single-channel pipettes:•2 mL• 2.5 mL•5 mL•10

Page 18

MaintenanceEppendorf Reference® 2English (EN)256.4.2.2 Opening and assembling the lower part (2 mL or larger) 1. : Unscrew the ejector sleeve.2. : K

Page 19 - 6.1 Cleaning

MaintenanceEppendorf Reference® 2English (EN)26The lower part consists of the following:Before assembling the pipette, first install the piston mount,

Page 20 - 6.2.1.3 After autoclaving

MaintenanceEppendorf Reference® 2English (EN)276.4.3.2 Removing and installing channels1. and : Position a pipette tip under the piston and careful

Page 21 - 6.3 Replacing O-rings

MaintenanceEppendorf Reference® 2English (EN)286.5 Adjusting pipettes6.5.1 General instructions on user and factory adjustmentThe Reference 2 was adju

Page 22 - PUSH TO RELEASE

MaintenanceEppendorf Reference® 2English (EN)29The existing dispensing volume must be checked before changing the factory calibration or adjustment. T

Page 23

Table of contentsEppendorf Reference® 2English (EN)3Table of contents1 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 24 - 6.4.2.1 Exchanging the filter

MaintenanceEppendorf Reference® 2English (EN)306.5.2 Changing the user adjustmentIf the adjustment is changed, the volume changes by a certain value.

Page 25

MaintenanceEppendorf Reference® 2English (EN)316.5.3 Changing the factory adjustmentAuxiliary aids from the delivery package• Red plastic safety plug•

Page 26

MaintenanceEppendorf Reference® 2English (EN)326.6 Decontamination before shipmentIf you wish to return the device to Eppendorf AG or an Eppendorf AG

Page 27

Technical dataEppendorf Reference® 2English (EN)337 Technical data7.1 Reference 2 Single-channel variableReference 2 single-channel variableModel Test

Page 28 - 6.5 Adjusting pipettes

Technical dataEppendorf Reference® 2English (EN)347.2 Reference 2 Multi-channel variableReference 2 single-channel variableModel Test tipepT.I.P.S. co

Page 29

Technical dataEppendorf Reference® 2English (EN)357.3 Reference 2 fixedReference 2 fixed volumeModel Test tip epT.I.P.S. color codeVolume rangeLengthE

Page 30 - %). If the volume

Technical dataEppendorf Reference® 2English (EN)367.4 Test conditions Test conditions and test evaluation match in compliance with ISO 8655, Part 6.

Page 31

Ordering InformationEppendorf Reference® 2English (EN)378 Ordering Information8.1 Single-channel pipettes, variable volume8.2 Single-channel pipettes,

Page 32

Ordering InformationEppendorf Reference® 2English (EN)388.3 Single-channel lower part for variable volume and fixed volume8.4 Multi-channel pipettesOr

Page 33 - 7 Technical data

Ordering InformationEppendorf Reference® 2English (EN)398.5 Multi-channel lower parts8.5.1 8-channel lower parts8.5.2 12-channel lower parts8.6 Spare

Page 34

Table of contentsEppendorf Reference® 2English (EN)46.5 Adjusting pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 35 - 7.3 Reference 2 fixed

Ordering InformationEppendorf Reference® 2English (EN)40

Page 36 - 7.5 Ambient conditions

PatentsEppendorf Reference® 2English (EN)419 PatentsU.S. Patent No. 7,434,484 Is used by Reference 2 pipettes with variable volume setting (single-cha

Page 37 - 8 Ordering Information

IndexEppendorf Reference® 2English (EN)42IndexAAssembly10 µL, 100 µL and 300 µL multi-channel lower part...

Page 39 - 8.5 Multi-channel lower parts

Evaluate your manualGive us your feedback.www.eppendorf.com/manualfeedbackYour local distributor: www.eppendorf.com/contactEppendorf AG · 22331 Hambur

Page 40 - English (EN)

Operating instructionsEppendorf Reference® 2English (EN)51 Operating instructions1.1 Using this manual Read this operating manual completely before u

Page 41 - 9 Patents

Operating instructionsEppendorf Reference® 2English (EN)6

Page 42

Product descriptionEppendorf Reference® 2English (EN)72 Product description2.1 FeaturesReference 2 pipettes are piston-stroke pipettes which operate a

Page 43

Product descriptionEppendorf Reference® 2English (EN)82.2 Delivery packageThe following accessories are included in the delivery package:Piece Descrip

Page 44 - Evaluate your manual

Product descriptionEppendorf Reference® 2English (EN)92.3 Main illustration2.3.1 Reference 2 single-channel1 Volume displayVolume display with four