Maintenance
Eppendorf Reference
®
2
English (EN)
24
6.4.2 Single-channel lower part
Applies to the following single-channel pipettes:
•2 mL
2.5 mL
•5 mL
•10 mL
Replace the protection filter in the tip cone each time it comes into contact with liquid.
6.4.2.1 Exchanging the filter
1. Remove the filter sleeve.
2. Use a pipette tip to push out the old filter.
3. Insert a new filter into the filter sleeve.