Page 1 - Centrifuge 5702 RH

Centrifuge 5702 / Centrifuge 5702 R /Centrifuge 5702 RH

Page 2 - 2012 Eppendorf AG, Hamburg

29The fixed-angle rotor F-35-30-17 can be loaded with 15 ml Falcons® and 15 ml round bottom centrifuge tubes. It is available in the 10 and 30 positio

Page 3 - Centrifuge 5702

30The swing-bucket rotor A-4-38 can be equipped with round or rectangular buckets. Round buckets and the associated adapters (see ordering information

Page 4 - Centrifuge 5702 R

31The inner part of the aerosol-tight cap is also equipped with a silicone sealing ring. Never remove this sealing ring from the cap. Whenever using t

Page 5 - Contents

32Note that the rcf (g-force) indicated when the display is switched is standardized to 15 ml Falcon® tubes in the rotor A-4-38. At 4,400 1/min you ca

Page 6 - 1 Introduction

33When working with Ficoll® density gradients or other sensitive applications (e.g. certain types of blood and urine centrifugations) slower accelerat

Page 7 - with a mild cleaning agent

34With the Centrifuge 5702 R / 5702 RH it is possible to store up to two permanent programs (only during standstill): First set the required program d

Page 8

35In the case of the Centrifuge 5702 RH, we are dealing with a benchtop centrifuge with a regulated heating and cooling system. The Centrifuge 5702 RH

Page 9

36The outer surfaces of the centrifuge and the rotor chamber should be cleaned regularly with a mild cleaning agent. This is for hygiene purposes and

Page 10 - 3 Operation

37When centrifuging glass tubes, please observe that high speeds/rcf’s increase the risk of glass breakage. Please follow the manufacturer’s instructi

Page 11

38If the suggested remedy repeatedly fails, please contact Service.Error Cause RemedyStand-by button lit red. Centrifuge not ready. Press STANDBY but

Page 12

No part of this publication may be reproduced without the prior permission of the copyright owner.Copyright© 2012 Eppendorf AG, Hamburg Task Lid Pres

Page 13

39Technical specifications subject to change!Line connection: Centrifuge 5702230 V / 50 – 60 Hz120 V / 50 – 60 Hz100 V / 50 – 60 HzCentrifuge 5702 R /

Page 14

40Centrifuge 5702230 V / 50 – 60 Hz, without rotor 5702 000.019Centrifuge 5702 R230 V / 50 – 60 Hz, without rotor 5703 000.012Centrifuge 5702 RH230 V

Page 15

41Adapter for standard tubes and blood sampling systems for round bucket 85 ml2 adapters 4 x 1.5 – 2.0 ml micro test tubes 5702 745.0062 adapters 5

Page 16

52La Centrifuge 5702 es una centrífuga de sobremesa no refrigerada, la Centrifuge 5702 R es una centrífuga de sobremesa refrigerada y la Centrifuge 57

Page 17 - 4 Maintenance and cleaning

53– Antes de proceder a la centrifugación, deben inspeccionarse siempre visualmente los recipientes para identificar posibles daños materiales. Los r

Page 18

54Después del centrifugado la tapa de la Centrífuga 5702 se abre automáticamente, mientras que la tapa de la Centrífuga 5702 R / 5702 RH refrigerada p

Page 19 - 5 Troubleshooting

55Desenchufe la clavija de la red de alimentación y esperar hasta que se pare el rotor (¡puede tardar hasta 5 min!). Desbloqueo de emergencia en la pa

Page 20 - 6 Technical data

56Si la medida propuesta no fuese efectiva repetidamente para la eliminación del fallo, por favor, póngase en contacto con el servicio técnico.Fallo C

Page 23 - Instrucciones breves

Abb. 1 / Fig. 11stand-byEspeedtimeshort23open1 Netzschalter und -stecker 1 Mains switch and plug2 Rotormutter 2 Rotor nut3Rotor 3RotorE Notentriegelun

Page 26

Your local distributor: www.eppendorf.com/contactEppendorf AG · 22331 Hamburg · [email protected] · www.eppendorf.com Eppendorf® and the

Page 27

Abb. 2 / Fig. 223openfast tempspeedstand-byprog 1 /2temptimeshortCE1Centrifuge 5702 RAbbildung 2 / Figure 21 Netzschalter und -stecker 1 Mains switch

Page 28

241 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 29

25Centrifuge 5702 is non-refrigerated, Centrifuge 5702 R is refrigerated and Centrifuge 5702 RH is a refrigerated and heatable bench-top centrifuge. F

Page 30

26For your personal safety, please be sure to comply with the following regulations unconditionally:– The centrifuge 5702 / 5702 R / 5702 RH must only

Page 31

27The following rotors and accessory buckets have a maximum service life of 7 years: A-4-38 5702 720.003 and 5702 740.004A-8-17 5702 700.002The date o

Page 32 - 5702 901.016-11/052013

28Refer to the frontal view on the first inside cover page of this manual.Fig. 1: Front view of the Centrifuge 5702.Fig. 2: Front view of the Centrifu