5.9.5 Loading a saved program
5.9.5.1 Loading program prog 1 to prog 5
On supply, the prog 1 to prog 5 program keys are assigned as follows:
1. In order to call up a program on the program spaces 1 to 5, press one of the program
keys prog 1 to prog 5.
The LED above the program key lights blue.
The display shows the parameters of the program.
2. In order to start the program, press the start/stop key.
5.9.5.2 Loading a program from the program list
1. In order to load a program from the program list, select the program under Menu >
Programs. Confirm with the menu/enter key.
2. Use the menu arrow keys to select the Load menu item. Confirm with the menu/enter
key.
The display shows the parameters of the program.
Programs with several program levels:
The display shows the parameters of the first program level. To display the
parameters of the other program levels, use the ◄ or ► menu arrow keys to select
the corresponding step.
3. In order to start the program, press the start/stop key.
The asterisk marks the active program level Step 1. The display shows the parameters
of Step 2.
Program name Temperature Mixing time/
temperature
control duration
Mixing
frequency
prog 1 key Cooling 8°C 0 rpm
prog 2 key Ligation 16°C 16:00 h 0 rpm
prog 3 key Restr. Digest 37°C 1:00 h 1 000 rpm
prog 4 key Prot. K Digest 56°C 10:00 min 1 000 rpm
prog 5 key Denaturation 95°C 30:00 min 0 rpm