Safety
Eppendorf ThermoMixer
®
C
English (EN)
3.4 Warnings for intended use
Read the operating instructions and observe the following general safety information
before using the ThermoMixer C.
DANGER! Risk of explosion.
Do not operate the device in areas where work is completed with explosive
substances.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device for processing any substances which could generate an
explosive atmosphere.
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning
or disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Use closed tubes and closed plates.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
WARNING! Electric shock due to damage to device or mains cable.
Only switch on the device if the device and mains cable are undamaged.
Only use devices that have been properly installed or repaired.
In case of danger, disconnect the device from the mains supply by pulling the
power plug from the device or the mains socket or, by using the isolating
device intended for this purpose (e.g., emergency stop switch in the
laboratory).
WARNING! Lethal voltages inside the device.
Touching parts which are under high voltage may cause an electric shock. An
electric shock injures the heart and causes respiratory paralysis.
Ensure that the housing is closed and undamaged.
Do not remove the housing.
Ensure that no liquid can penetrate into the device.
Only authorized service staff may open the device.