Eppendorf Multipette Xstream User Manual Page 42

  • Download
  • Add to my manuals
  • Print
  • Page
    / 58
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 41
42
Technical data
Multipette
®
/Repeater
®
(X)stream
English (EN)
8 Technical data
8.1 Errors with the Multipette (X)stream/Repeater (X)stream
Combitip advanced Volume range Testing
volume
Error limits
Error
Systematic error Random error
± % ± µL ± % ± µL
0.1 mL
white
Increment:
0.1 µL
1 µL – 100 µL 10 µL 1.60 0.16 2.50 0.25
50 µL 1.00 0.50 1.50 0.75
100 µL 1.00 1.00 0.50 0.50
0.2 mL
light blue
Increment:
0.2 µL
2 µL – 200 µL 20 µL 1.30 0.26 1.50 0.30
100 µL 1.00 1.00 1.00 1.00
200 µL 1.00 2.00 0.50 1.00
0.5 mL
purple
Increment:
0.5 µL
5 µL – 500 µL 50 µL 0.90 0.45 0.80 0.40
250 µL 0.90 2.25 0.50 1.25
500 µL 0.90 4.50 0.30 1.50
1 mL
yellow
Increment:
1 µL
10 µL – 1000 µL 100 µL 0.90 0.90 0.55 0.55
500 µL 0.60 3.00 0.30 1.50
1 000 µL 0.60 6.00 0.20 2.00
2.5 mL
green
Increment:
2.5 µL
25 µL – 2 500 µL 250 µL 0.80 2.00 0.45 1.125
1 250 µL 0.50 6.25 0.30 3.75
2 500 µL 0.50 12.5 0.15 3.75
5 mL
blue
Increment:
5 µL
50 µL – 5 000 µL 500 µL 0.80 4.00 0.35 1.75
2 500 µL 0.50 12.50 0.25 6.25
5 000 µL 0.50 25.00 0.15 7.50
Page view 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 57 58

Comments to this Manuals