Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)102.5 epT.I.P.S. longIn addition to the "standard tips", a second longer tip is
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)112.6 AltitudeAt increased altitudes there is a fall in air pressure so that the aspirati
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)122.7.1 1-point adjust.After you enter the density, the selected volume and the correspon
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)13Abb. 2-1: Example: Correction of the piston stroke with "Abb. 2-2: 1-point adjust.
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)14Abb. 2-4: Example: Changing the piston stroke of a mechanical pipetteFig. 2-4: Example:
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)152.7.2 2-point adjust.After you enter the density, two different volumes and the corresp
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)162.7.3 3-point adjust.After you enter the density, three different volumes and the mean
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)172.8 Marking the Xplorer pipette to show the adjustment has been changedYou must carry o
ErrorEppendorf Xplorer®/Xplorer® plusEnglish (EN)183 Error3.1 Limiting values for measurement deviations according to ISO 8655 ddd3.1.1 Xplorer pipett
ErrorEppendorf Xplorer®/Xplorer® plusEnglish (EN)193.1.2 Xplorer pipette, multi-channelModel Test tipepT.I.P.S. Color codeVolume rangeLengthTesting vo
Adjustment - en-02/042013 Copyright© 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the cop
ErrorEppendorf Xplorer®/Xplorer® plusEnglish (EN)203.2 Eppendorf AG limiting values for measurement deviations3.2.1 Xplorer pipette, single-channelpip
ErrorEppendorf Xplorer®/Xplorer® plusEnglish (EN)213.2.2 Xplorer pipette, multi-channelpipetteIncrementPipette tipColor codeVolume rangeLengthTesting
Speed table Eppendorf XplorerEppendorf Xplorer®/Xplorer® plusEnglish (EN)224 Speed table Eppendorf Xplorer4.1 Speed table for Xplorer single-channelTh
Volume limitEppendorf Xplorer®/Xplorer® plusEnglish (EN)235 Volume limit6 Additional volumes during aspirationep Dualfilter T.I.P.S. Volume limitAds/D
PatentsEppendorf Xplorer®/Xplorer® plusEnglish (EN)247 PatentsU.S. Patent No. 6,499,365 Is used by Xplorer / Xplorer plus pipettes (single-channel and
Evaluate your manualGive us your feedback.www.eppendorf.com/manualfeedbackYour local distributor: www.eppendorf.com/contactEppendorf AG · 22331 Hambur
Table of contentsEppendorf Xplorer®/Xplorer® plusEnglish (EN)3Table of contents1 General information on adjustment . . . . . . . . . . . . . . . . .
Table of contentsEppendorf Xplorer®/Xplorer® plusEnglish (EN)4
General information on adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)51 General information on adjustmentThe piston stroke of Xplorer pipettes
General information on adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)6Before the adjustment is changed, the existing adjustment must be checke
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)7If you would like to conduct a 1-3 point adjustment for a liquid other than water on th
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)82.1 Pswrd. protectionFor the Xplorer plus, the adjustment can be protected with a four-d
Change adjustmentEppendorf Xplorer®/Xplorer® plusEnglish (EN)92.2.1 Reset to the factory setting1. Move the selection dial to Opt.2. Select the Adjust