Universal Stand for mounting Eppendorf Micromanipulators
InjectMan NI 2, TransferMan NK 2, PatchMan NP 2
next to the microscope
To guarantee its stability, the Universal Stand requires a ferromagnetic ground plate. The
magnetic force stabilizes the positioning of the Universal Stand. This plate does not form part of
the delivery package. It can be obtained via specialist distributors or from a suitable retail outlet.
The plate should be a least 4 mm thick. The size and the weight should be sufficient to ensure
that the weight of the motor module can be borne safely (e.g. 8 x 300 x 600 mm, approx. 13 kg).