Operating manualDraft — Biological Shakers Innova® 44/44Rseeon p.Fig.Tab.p.Operating manualNew Brunswick™ Biological Shakers Innova® 44/
Biological Shakers Innova® 44/44R — Operating manual102.5.2 Incorrect handling of accessoriesCAUTION!Lack of safety due to incorrect accessories or
11Biological Shakers Innova® 44/44R — Operating manual3 Product descript ion3 Product description3.1 Main illustrationAbb. 1: Front and r ear view o
Biological Shakers Innova® 44/44R — Operating manual12Abb. 2: Internal viewFig. 2: Internal view3.2 General overviewThe Innova 44/44R stackable incu
13Biological Shakers Innova® 44/44R — Operating manualThe Innova 44/44R is equipped with visual and/or audible alarms that alert the user to the fol
Biological Shakers Innova® 44/44R — Operating manual144 Inspection a nd unpacking o f equipment4 Inspection and unpacking of equipment4.1 Inspection
15Biological Shakers Innova® 44/44R — Operating manual5 Preparing the location5 Preparing the location5.1 Physical locationIt is essential that the
Biological Shakers Innova® 44/44R — Operating manual16Abb. 3: Innova 44/44R Front View Two Shakers Stacked on 12 in BaseFig. 3: Innova 44/44R Front
17Biological Shakers Innova® 44/44R — Operating manual6 Installation6 Installation6.1 Tools required for installationTo install the Innova 44/44R on
Biological Shakers Innova® 44/44R — Operating manual181. Immobilize the top locknut against the unit with one wrench whenever you adjust the foot, t
19Biological Shakers Innova® 44/44R — Operating manualAbb. 6: Mount Sh aker on B aseFig. 6: Mount Shaker on Base5. Using the 1/2 – 13 x 1 1/4 in scr
M1282-0050-N/022013Copyright © 2013 Eppendorf AG, Germany. No part of this publication may be reproduced without the prior permission of the copyrig
Biological Shakers Innova® 44/44R — Operating manual20Abb. 7: Shaker Installe d on BaseFig. 7: Shaker Installed on Base10. Make sure the shaker is l
21Biological Shakers Innova® 44/44R — Operating manualAbb. 8: Install I nnova 44/44 R Stacking KitFig. 8: Install Innova 44/44R Stacking Kit4. Usin
Biological Shakers Innova® 44/44R — Operating manual22Abb. 9: Install 4 400/4430 t o 44/44R Stacking KitFig. 9: Install 4400/4430 to 44/44R Stacki
23Biological Shakers Innova® 44/44R — Operating manual3. With two assistants guiding the unit from opposite sides, lower the shaker onto the mountin
Biological Shakers Innova® 44/44R — Operating manual246.8 Stack a 4400/4430 on a 44/44R1. Using a forklift or lifter, raise the Innova 4400/4430 to
25Biological Shakers Innova® 44/44R — Operating manual7Features7Features7.1 ControlsAbb. 11: Front Panel ( detail)Fig. 11: Front Panel (detail)• STA
Biological Shakers Innova® 44/44R — Operating manual26Abb. 12: Control Panel (Righ t Side)Fig. 12: Control Panel (Right Side)7.2 LCD displayWhen th
27Biological Shakers Innova® 44/44R — Operating manualAbb. 13: Display Scree nFig. 13: Display Screen• For an explanation of the icons on the displa
Biological Shakers Innova® 44/44R — Operating manual28Tab. 1: Screens1 See table below, Display Screen Parameters; 2 Optional; 3 Not installedScreen
29Biological Shakers Innova® 44/44R — Operating manualTab. 2: Display Screen Parameters1 Optional7.4 Display iconsThere are six icons that help iden
3Biological Shakers Innova® 44/44R — Operating manual1 Table of contentsTable of contents1 User instructions . . . . . . . . . . . . . . . . . . . .
Biological Shakers Innova® 44/44R — Operating manual30Tab. 3: Display Icons7.5 AlarmsIf an alarm condition exists, the field in the lower right corn
31Biological Shakers Innova® 44/44R — Operating manual7.6 Glide-up doorThe Innova 44/44R is equipped with a space-saving glide-up door. When the doo
Biological Shakers Innova® 44/44R — Operating manual328 Getting started8 Getting started8.1 Platform assembliesThe Innova 44/44R can be used with a
33Biological Shakers Innova® 44/44R — Operating manual1. With reference to figure below (see Fig. 15 on p. 33), place the platform on the subplatfor
Biological Shakers Innova® 44/44R — Operating manual344. Place the second girdle around the upper portion of clamp body (just as the first girdle wa
35Biological Shakers Innova® 44/44R — Operating manualAbb. 18: Rear PanelFig. 18: Rear Panel8.5 Optional gassing manifold kitThis option is factory-
Biological Shakers Innova® 44/44R — Operating manual36Abb. 19: Front GrilleFig. 19: Front Grille2. Check the drain hose to be sure it is clamped tig
37Biological Shakers Innova® 44/44R — Operating manual9 Operation9 Operation9.1 Safety precautionsBefore operating the shaker, verify that anyone in
Biological Shakers Innova® 44/44R — Operating manual38Abb. 20: Chang ing Disp lay ParameterFig. 20: Changing Display Parameter2. Click the Control K
39Biological Shakers Innova® 44/44R — Operating manualAbb. 22: Changing SetpointFig. 22: Changing Setpoint3. Turn or spin the Knob to reset the setp
Biological Shakers Innova® 44/44R — Operating manual47.3 Changing screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Biological Shakers Innova® 44/44R — Operating manual40Abb. 23: Summar y ScreenFig. 23: Summary ScreenThe only elements you can modify in this screen
41Biological Shakers Innova® 44/44R — Operating manualAbb. 24: Setup ScreenFig. 24: Setup ScreenTo change the day:1. Turn the Knob to highlight the
Biological Shakers Innova® 44/44R — Operating manual42To move out of this screen and into the next:1. Use the control Knob to highlight SET, then cl
43Biological Shakers Innova® 44/44R — Operating manual9.4.5 RS232 ScreenThis screen (see Fig. 26 on p. 43) is used only if you have connected a pers
Biological Shakers Innova® 44/44R — Operating manual44Tab. 4: Communication ModeTo change the Baud Rate:1. Turn the control knob to highlight the cu
45Biological Shakers Innova® 44/44R — Operating manualAbb. 28: Program s ScreenFig. 28: Programs Screen9.5 Program the shaker9.5.1 Timer onlyBy sett
Biological Shakers Innova® 44/44R — Operating manual462. Turn the Knob until the field says New. Click the Knob in to select this mode. The screen f
47Biological Shakers Innova® 44/44R — Operating manualset, use the Control Knob to select the field, click the Knob to select the field (it will fla
Biological Shakers Innova® 44/44R — Operating manual48The three-step program we have illustrated is designed to start incubating cultures at 37 °C a
49Biological Shakers Innova® 44/44R — Operating manualTo stop a program: You can abort a run any time by changing the program’s function to Off.9.6
5Biological Shakers Innova® 44/44R — Operating manual11.2 Cleaning external and internal surfaces . . . . . . . . . . . . . . . . . . . . . . . . .
Biological Shakers Innova® 44/44R — Operating manual50Abb. 34: Calibrate S creenFig. 34: Calibrate Screen9.8 Using CalspeedThe Calspeed function, wh
51Biological Shakers Innova® 44/44R — Operating manual3. Pull both side levers (see Fig. 35 on p. 51) down. This will disengage the platform from th
Biological Shakers Innova® 44/44R — Operating manual5210 Troubleshooting10 Troubleshooting10.1 General troubleshootingIf any problems occur with you
53Biological Shakers Innova® 44/44R — Operating manual Shaker does not run at set speed. Shaker is running in Program mode; Look for Run icon in dis
Biological Shakers Innova® 44/44R — Operating manual54 Photosynthetic GRO lamp does not work. Shaker is running in Program mode; Look for Run icon i
55Biological Shakers Innova® 44/44R — Operating manual11 Mainten ance11 Maintenance11.1 Routine maintenanceNo routine maintenance schedule is requir
Biological Shakers Innova® 44/44R — Operating manual5612 Technical da ta12 Technical data12.1 SpecificationsThese specifications assume a maximum lo
57Biological Shakers Innova® 44/44R — Operating manual1 Use of baffled flasks will significantly reduce maximum speed for any shaker.2 depending on
Biological Shakers Innova® 44/44R — Operating manual5813 Ordering infor mation13 Ordering information13.1 Replacement partsIn the table below you wi
59Biological Shakers Innova® 44/44R — Operating manual1 Clamps for Universal Platform are sold separately.Flask TypeCapacity150 ml Erlenmeyer Flasks
Biological Shakers Innova® 44/44R — Operating manual61Table of contents
Biological Shakers Innova® 44/44R — Operating manual60Tab. 5: Platform Capacity Chart1Maximum flask size when using Photosynthetic Light Bank is 4 L
61Biological Shakers Innova® 44/44R — Operating manual13.2.2 Flask clamps for universal platformsThe following clamps, according to flask size, are
Biological Shakers Innova® 44/44R — Operating manual621 Universal Platform Required13.2.5 Optional bases13.2.6 Optional stacking kit13.2.7 Optional
63Biological Shakers Innova® 44/44R — Operating manual14 Transport, storage and disposal14 Transport, storage and disposal14.1 Transport and storage
Biological Shakers Innova® 44/44R — Operating manual6415 Appendix A: Remote prog ramming15 Appendix A: Remote programming15.1 Setting up with HyperT
65Biological Shakers Innova® 44/44R — Operating manualTab. 6: COM Parameters for HyperTerminalSample Test:1. To request the current setpoint, type R
Biological Shakers Innova® 44/44R — Operating manual66If a command has additional arguments, the ASCII space control character separates each argume
67Biological Shakers Innova® 44/44R — Operating manual15.5 Profile control commandsFor example, the following PW command:PW<Space>25.0<Spac
Biological Shakers Innova® 44/44R — Operating manual6815.6 Report request commands15.7 Set/Get date and timeFor example, this =D command:=D<Space
69Biological Shakers Innova® 44/44R — Operating manual16 Appendix B: Certifications16 Appendix B: Certifications16.1 CertificationsThe Innova 44 an
7Biological Shakers Innova® 44/44R — Operating manual1 User in struction s1 User instructions1.1 Using this manualCarefully read this operating man
Biological Shakers Innova® 44/44R — Operating manual70IndexIndexAAccessories ...58
71Biological Shakers Innova® 44/44R — Operating manualReplacement parts...58Routine maintena
New Brunswick Scientific • www.nbsc.com • [email protected] • For your local sales office or distributor contact: Eppendorf AG · 22331 Hamburg · Germ
Biological Shakers Innova® 44/44R — Operating manual82Safety2 Safety2.1 Intended useThis device is exclusively intended for indoor use and for unifo
9Biological Shakers Innova® 44/44R — Operating manual2.5.1 Personal injury and damage to device WARNING!Electric shock due to damage to the device o