Temperatursensor für 384er BlockTemperature sensor for the 384well blockBedienungsanleitung · Operating ManualTS384_Titel_Rueck.fm Seite 1 Mittwoch,
22Protocol sheet for temperature validation (template)Mastercycler type: Device no: _________________________ _________________________(Date) (Signa
234.2.2 Validation of the Mastercycler 384 with software version 2.30 and aboveValidate the Mastercycler 384 as described in the operating manual for
244.3.3 Calibration / adjustment of the Mastercycler pro and Mastercycler ep 384 using the CycleManager pro / proXL or ep CycleManager Cal
25The temperature sensor does not require regular maintenance. Soiling can be carefully wiped off with a damp cloth. The oil film on the temperature se
26Sensor: Temperature sensor for the 384well blockThe operating conditions and the storage and transport conditions are included in the operating man
Eppendorf South Pacific Pty. Ltd.Tel. +61 2 98 89 50 00Fax +61 2 98 89 51 11E-Mail: [email protected]: www.eppendorf.com.auEppendorf Austri
In touch with lifeYour local distributor: www.eppendorf.com/worldwideEppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 538 01-0 · Fax: +49 40 538 0
2Bedienungsanleitung. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Operating Ma
151 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161.1 How to u
16The 348-well block sensor and the temperature validation system enable validation and calibration / adjustment of the Mastercycler 384, Mastercycler
17– Temperature sensor for the 384well block– Operating manual for the 384 temperature sensor (this manual)– Silicone mat– Personal card for use with
184.1.1 Filling the measurement positions with oilTo ensure good heat transfer from the wall of the measurement positions to the temperature sensor,
19Before insertion of the sensor, the four spacers are to be placed into positions B5, B20, O5 and O20. Insert the sensor into the indicated position
20The validation program tests the accuracy of the temperature of the block. The program procedure selects consecutive temperatures. The temperatures
214.2.1 Validation of the Mastercycler 384 with software version 2.22Validation of the Mastercycler 384 with the software version 2.22 is carried out
Comments to this Manuals